Measuring Tips

Turn the meter on and off with the

I/O/EXIT

key.

To review a calibration, press the

REVIEW

key.

When using the dissolved oxygen probe, polarize the probe by
plugging it into the meter before attempting a calibration or
measurement. The length of time required for polarization depends
on how long the electrode has been unplugged. It can vary from 60
minutes (for a new probe) to a few seconds (if you momentarily
disconnect the probe). If the

CAL

and

?

icons flash after calibrating

the DO probe, the probe is not polarized.

Connect the pH electrode to the black connector, and the DO or
conductivity electrodes to the blue connector.

If the meter is not in reading mode, it may take a few key presses to
turn the meter off.

If the readings do not change when you move an electrode to a new
sample, the meter probably has the Display Lock set to

ON

for the

parameter you are reading. If the

Lock Icon

appears in the lower left

portion of the display, the setting is on. To change the setting, use
setup number 5. Or, press

READ

to initiate a new reading.

When calibrating pH or dissolved oxygen, make sure that the
instrument is displaying readings to the correct number of decimal
places for your needs. This will assure that your calibration is
sufficiently precise. To change the setting, use setup number 6.

© Hach Company, 2000. All rights reserved. Printed in the U.S.A.

ap/dk 1/00 1ed

Содержание SensION156

Страница 1: ...54650 23 and Multiparameter Meters ...

Страница 2: ...esired option is selected press EXIT to return to the reading mode Setup 5 Display Lock From pH reading mode press SETUP Press the UP ARROW until the Setup number is 5 Press ENTER to toggle display lock off and on When the desired option is selected press exit to return to the reading mode Setup 6 Measurement Resolution From pH reading mode press SETUP Press the UP ARROW until the Setup number is ...

Страница 3: ... by three beeps If measuring in mV mode the three beeps will still sound when the stabilization occurs but ENTER must be pressed to accept the reading This lets the operator control the acceptance point of the buffer 5 When the reading has stabilized or been accepted the standard number will change to 2 6 Remove the probe from the first buffer and rinse with deionized water Place the probe into th...

Страница 4: ...onductivity reading mode press SETUP Press the UP ARROW until the Setup number is 7 Press ENTER Press the UP or DOWN ARROW keys to switch between linear and non linear correction functions To choose a non linear conversion scroll until a flashing NaCl appears then press ENTER To select a linear conversion scroll until the conversion coefficient appears Use the numeric keypad to set the value of th...

Страница 5: ... the most recent type of calibration 5 Look at the units field to see what kind of calibration is active Scroll to the preferred units using the UP or DOWN ARROWS 6 Use the number keys to change the numeric value if desired It is not necessary to fill up the numeric entry screen before moving on To clear the numeric display press CE 7 When the value and units are correct press ENTER to calibrate o...

Страница 6: ...press EXIT to return to the reading mode Setup 7 Salinity Factor Determine sample salinity The units for salinity are parts per thousand 0 00 From dissolved oxygen reading mode press SETUP Press the UP ARROW until the Setup number is 7 Press ENTER Use the number keys to enter a salinity factor ranging from 0 to 42 Press ENTER to accept the value or EXIT to leave the value unchanged When the desire...

Страница 7: ...n the chamber to reach a steady state Note Gently squeezing the lower chamber a couple of times to force water saturated air into the probe chamber will speed up stabilization Note Keep the DO probe at a uniform temperature When holding the probe do not touch the metallic button on the side of the probe The button is a thermistor that senses temperature An inaccurate calibration will result if the...

Страница 8: ...rily disconnect the probe If the CAL and icons flash after calibrating the DO probe the probe is not polarized Connect the pH electrode to the black connector and the DO or conductivity electrodes to the blue connector If the meter is not in reading mode it may take a few key presses to turn the meter off If the readings do not change when you move an electrode to a new sample the meter probably h...

Отзывы: