Troubleshooting
Software and/or Hardware
related
The computer shows that
an Application Error has
occurred.
Application errors can occur when the computer is asked to do
something it does not understand or it cannot do. These errors
are more frequently found when more than one program is
running on the computer at the same time. The system will
typically offer two choices: Retry or Ignore. Select Ignore to
close the application or program where the problem occurred.
Exit Windows
™
and restart that application again.
It is important to identify whether the error is reproducible or
random. Document a random error for future reference. If the
problem can be reproduced, contact the manufacturer
for assistance.
Chemistry/Flow/Hardware/
Software related
Check standards keep
failing.
There are many causes for a check standard to fail. Repeating
the run over and over again will not fix the problem. Look for the
causes of the problem. Are the check standards made properly?
Are there flow problems? Make sure the calibration curve is
acceptable. The “R” should be 0.995 or better. In general,
residuals should be less than 1% for the higher standards and
less than 10% for the lower standards.
Software and/or Hardware
related
Sampler time out
The sample line or dilutor tubing may impede the free movement
of the sample probe. Make sure that the system has the latest
version of software installed. A loose or unplugged sampler
cable may be responsible for sampler time outs. Refer to
for correct assembly.
Software related
The heater element does
not heat.
Make sure the heater temperature has been set for that channel
in Run Properties>Analytes.
Chemistry/Flow/Hardware/
Software related
Calibration fails
There are many causes for a calibration to fail. Repeating the
run repeatedly will not fix the problem. Look for the causes of the
problem. Are the check standards made properly? Are there flow
problems? Are the peaks being integrated properly? Are the
specified concentrations correct. Are there peaks? Is the
response very low?
Software related
The baseline is not at
zero.
This is usually not a problem. The baseline will most likely be at
a value other than zero depending on the color of the reagents,
interference filter, etc. The baseline could be much higher than
normal for a specific analysis based on historic data. This could
be due to reagent contamination or old reagents. Reagent
contamination and reagents that get darker as they get older will
affect the analysis.
Note:
Make sure the baseline is close to zero when running DI
water or air through the flow cell. If not, contact Technical
Support.
Software related
Probe keeps hitting cups.
Make sure the probe is not bent. Seat the sample racks
correctly. Align the probe with the wash bath. Contact the
manufacturer if this problem persists.
Software related
Peaks are smaller than
the ones obtained in the
past.
Check the manifold against the manifold diagram in the method.
Make sure that the reagent lines are in the correct reagent
containers. Verify the sample loop. Check the standards
preparation procedure. Make sure the reagents are made
properly. Test the Valve Timing.
Table 14 Troubleshooting problems (continued)
Problem type
Problem description
Solution
Содержание QC8500 Series 2
Страница 2: ......
Страница 6: ...4 Table of Contents ...
Страница 10: ...8 Specifications ...
Страница 30: ...28 Installation ...
Страница 52: ...50 Operation ...
Страница 84: ...82 Troubleshooting ...
Страница 88: ...86 Replacement Parts and Accessories ...
Страница 90: ...88 ...
Страница 92: ...90 Warranty ...
Страница 94: ...92 Certification ...
Страница 102: ...100 Ion Chromatography ...