15
Installation
3.
Place the glass weight end of the waste line in a proper waste container. Never
submerge the waste line in the waste solution because this may cause back pressure
and flow restrictions. Make sure the waste solution always drips into the container.
3.1.6 Manifold installation
The manifold is the reaction module where the chemistry occurs.
Specific installation requirements are detailed in the
included with
each chemistry method. Refer to
and the following steps to install
the manifold for flow injection analysis.
To install the manifold:
1.
On the valve, connect port 1 to port 4 with the sample loop.
2.
Connect the manifold outlet to the flow cell inlet (
3.
Install the interference filter (
, item 5).
4.
Connect the carrier line to port 2 on the valve (
5.
Connect the tubing from port 3 to the connector labeled From Valve.
6.
Attach one end of a waste line to the flow cell outlet. Drop the other end into a waste
container (
, item 4).
Figure 10 Flow injection connections
1
System unit
5
Waste lines
2
Channel 1
6
Waste container
3
Channel 2
7
Waste line
4
Channel 3
8
Sample line
Содержание QC8500 Series 2
Страница 2: ......
Страница 6: ...4 Table of Contents ...
Страница 10: ...8 Specifications ...
Страница 30: ...28 Installation ...
Страница 52: ...50 Operation ...
Страница 84: ...82 Troubleshooting ...
Страница 88: ...86 Replacement Parts and Accessories ...
Страница 90: ...88 ...
Страница 92: ...90 Warranty ...
Страница 94: ...92 Certification ...
Страница 102: ...100 Ion Chromatography ...