60
Maintenance
5.12 Manifold maintenance
Follow the recommended cleaning procedures listed in the QuikChem methods. In
addition, every 500 hours of use, clean each fitting and replace the O-rings as necessary.
If any of the tubing is stained, replace it with new tubing. Replace all manifold tubing once
a year.
5.13 Computer maintenance
Refer to the computer manufacturer manual for maintenance requirements.
5.14
Lamp replacement
DANGER
Potential Electrocution Hazard. Disconnect AC input before opening any part of the
System Unit and before removing or replacing the lamp. Make all wiring
connections and replace and secure all chassis parts before reconnecting AC
input.
CAUTION
Potential Burn Hazard. The lamp is hot when the System Unit is turned on. Turn the
System Unit off and allow 30 minutes for the lamp to fully cool before removing.
After installing a new lamp, make sure all wiring is dressed away from hot
surfaces.
Note:
Handle the lamp on the outside of the reflector and base only. Do not handle lamp bulb glass.
for lamp replacement. Do not touch the lamp bulb
during installation.
Figure 51 Detector module connection
1
Detector module (bottom view)
3
Detector cable
2
Set screws for fiber optic cable
4
Fiber optic cable
Содержание QC8500 Series 2
Страница 2: ......
Страница 6: ...4 Table of Contents ...
Страница 10: ...8 Specifications ...
Страница 30: ...28 Installation ...
Страница 52: ...50 Operation ...
Страница 84: ...82 Troubleshooting ...
Страница 88: ...86 Replacement Parts and Accessories ...
Страница 90: ...88 ...
Страница 92: ...90 Warranty ...
Страница 94: ...92 Certification ...
Страница 102: ...100 Ion Chromatography ...