Table 7 Warning list for conductivity sensors (continued)
Warning
Description
Resolution
CAL OVERDUE
The Cal Reminder time
has expired
Calibrate the sensor.
NOT
CALIBRATED
The sensor has not been
calibrated
Calibrate the sensor.
REPLACE
SENSOR
The sensor has been in
operation > 365 days
Calibrate the sensor with a
reference solution and reset
the sensor days. Refer to
Sensor diagnostic and test
menu
on page 16. If the
calibration fails, call technical
support.
CAL IN
PROGRESS
A calibration was started
but not completed
Return to calibration.
OUTPUTS ON
HOLD
During calibration, the
outputs were set to hold for
a selected time.
The outputs will become active
after the selected time period.
WRONG LINEAR
TC
The user-defined linear
temperature compensation
is out of range
The value must be between
0 and 4%/°C; 0 to 200 °C.
WRONG TC
TABLE
The user-defined
temperature compensation
table is out of range
The temperature is above or
below the temperature range
defined by the table.
Event list for sensors
The event list shows current activities such as configuration changes,
alarms, warning conditions, etc. To show the events, press the
MENU
key and select Sensor Diag, [Select Sensor], Event List. A list of possible
events is shown in
. Previous events are recorded in the event
log, which can be downloaded from the controller. Refer to the controller
documentation for data retrieval options.
Table 8 Event list for conductivity sensors
Event
Description
CAL READY
The sensor is ready for calibration
CAL OK
The current calibration is good
TIME EXPIRED
The stabilization time during calibration expired
CAL FAIL
The calibration failed
CAL HIGH
The calibration value is above the upper limit
K OUTRANGE
The cell constant K is out of range for the current
calibration
UNSTABLE
The reading during calibration was unstable
CHANGE IN CONFIG float The configuration was changed—floating point type
CHANGE IN CONFIG text
The configuration was changed—text type
CHANGE IN CONFIG int
The configuration was changed—integer value type
RESET CONFIG
The configuration was reset to the default options
POWER ON EVENT
The power was turned on
ADC FAILURE
The ADC conversion failed (hardware failure)
FLASH ERASE
The external serial flash memory erase occurred
TEMPERATURE
The temperature is out of range (-20 to 200 °C)
SAMPLE CAL START
Start of calibration for conductivity
SAMPLE CAL END
End of calibration for conductivity
ZERO CAL START
Start of zero calibration
ZERO CAL END
End of zero calibration
18
English
Содержание Polymetron
Страница 2: ...English 3 Français 20 Español 38 Português 58 中文 76 日本語 92 한글 109 ไทย 125 2 ...
Страница 8: ...1 2 3 4 8 English ...
Страница 9: ...5 6 7 8 English 9 ...
Страница 25: ...1 2 3 4 Français 25 ...
Страница 26: ...5 6 7 8 26 Français ...
Страница 42: ...Figura 3 Ejemplos de montaje 1 3 4 pulg NPT 2 1 4 pulg NPT 42 Español ...
Страница 44: ...1 2 3 4 44 Español ...
Страница 45: ...5 6 7 8 Español 45 ...
Страница 63: ...1 2 3 4 Português 63 ...
Страница 64: ...5 6 7 8 64 Português ...
Страница 79: ...图 3 安装示例 1 流通室 T 型安装 聚氯乙烯 英寸 NPT 2 流通室 T 型安装 不锈钢 1 4 英寸 NPT 中文 79 ...
Страница 81: ...1 2 3 4 中文 81 ...
Страница 82: ...5 6 7 8 82 中文 ...
Страница 95: ...図 3 取り付け例 1 フロースルー T 字取り付け PVC 19 mmNPT 2 フロースルー T 字取り付け ステンレス鋼 6 mmNPT 日本語 95 ...
Страница 97: ...1 2 3 4 日本語 97 ...
Страница 98: ...5 6 7 8 98 日本語 ...
Страница 112: ...그림 3 장착 예 1 흐름 통과 T 장착부 PVC 3 4 인치 NPT 2 흐름 통과 T 장착부 스테인리스 스틸 1 4 인치 NPT 112 한글 ...
Страница 114: ...1 2 3 4 114 한글 ...
Страница 115: ...5 6 7 8 한글 115 ...
Страница 130: ...1 2 3 4 130 ไทย ...
Страница 131: ...5 6 7 8 ไทย 131 ...
Страница 142: ......
Страница 143: ......