31
pH, High Purity Water,
continued
13.
The display will
show
Store?
.
Press
ENTER
to store the
calibration or
EXIT
to
leave the calibration
mode without storing
the values.
14.
Condition the
electrode for 10
minutes in deionized
water.
15.
Remove from
deionized water. Prime
the reference junction
by clicking the pump
until fresh gel emerges.
16.
Rinse the
electrode with
deionized water or
sample and blot dry.
17.
Place the
electrode into the
clean LIS Sample
Chamber Lid.
Note: Make sure that
the electrode will be
completely submerged
in the sample.
18.
Screw the cap and
attached electrode
onto a clean sample
bottle.
Note: EPA certified
clean bottles are
available. See
OPTIONAL
APPARATUS on
page 34.
READ
enter
OR
FILL
LINE