50
Troubleshooting,
continued
Out-of-
range
response
See “No response”, above.
The bulb may be broken.
Erratic
response
Reference junction contains
bubbles.
Push dispenser button and
turn clockwise two turns.
Electrode is not plugged in
to selected meter channel.
Plug electrode into meter
channel.
Reference junction has not
been properly primed with
gel.
Follow instructions in
.
Extreme external electrical
fields are present. (The pH
cable, though shielded, can
act as an antenna.)
Install a shorting cap on
the unused meter terminal.
Do not use stirrers in LIS
solutions.
Avoid static electricity,
nearby pumps, power
supplies, etc.
pH sensing bulb and
reference junction are not in
contact with the sample.
Submerge the probe
deeper into the sample.
The sample chemistry is
changing.
None.
The pH of a LIS solution is
changing due to CO
2
uptake.
Use LIS Chamber
Accessory.
The bulb is cracked,
discolored, or contaminated.
If dirty, clean according to
.
If cracked, discard.
The sample container is
contaminated.
Use a clean sample
container.
Symptom
Cause
Remedy