51
Troubleshooting,
continued
Low slope
Bulb is old.
Replace electrode.
Bulb is dirty.
Clean according to
Section
Standards are inaccurate.
Make sure standards were
entered into meter
correctly.
Use only NIST traceable
buffers.
Use fresh buffers.
Change buffers frequently.
History of harsh usage.
Replace electrode.
Cracked bulb.
Replace electrode.
High slope
Buffers are incorrect.
Use appropriate buffers.
Incorrect measurement or
calibration technique.
Follow procedures
described in
.
Temperature variations in
standards.
Keep sample and standards
at same temperature.
Standards contaminated.
Prepare fresh standards.
High offset
Leakage pathway results in
stray voltage via internal
electrical short.
Return electrode under
warranty.
Reference gel has become
diluted or junction is
plugged by solidified gel or
an air bubble.
Replace cartridge and
reprime.
Cracked bulb.
Replace electrode.
A 5-pin electrode that is not
in use is connected to the 5-
pin connector.
Disconnect the 5-pin
electrode.
Sluggish
response
Sample is cold and of low
ionic strength.
Wait patiently. Samples
may respond slowly.
Improper conditioning.
Condition according to
Section
Bulb, guard, and electrode
stem are dirty.
Clean according to
Section
Symptom
Cause
Remedy
Содержание PLATINUM Series
Страница 2: ...2...
Страница 4: ...4...
Страница 6: ...6...
Страница 8: ...8...
Страница 16: ...16...
Страница 17: ...17 Section 2 Applications...
Страница 18: ...18...
Страница 26: ...26...
Страница 42: ...42...
Страница 46: ...46...
Страница 52: ...52...
Страница 54: ...54...
Страница 60: ...60...
Страница 64: ......