pH, Low Ionic Strength (LIS) Water,
continued
10.
When a stable pH
is determined, the
display will show
Standard 2?
.
Remove the electrode
from the cup. Rinse
with deionized water
and blot dry.
Note:
The sens
ion
™4 meter
will select a stable
reading using default
meter parameters and
the specified resolution.
To override the default
meter parameters, see
the meter manual.
11.
Place the electrode
in the 4.0 (or 10.0) pH
buffer. Press
ENTER
.
12.
After the last
calibration point has
stabilized and the
display reads
Standard 3?,
press
EXIT
.
Note: For a three-point
calibration, repeat steps
10-11 with an additional
buffer.
13.
The display will
show
Store?
. Press
ENTER
to store the
calibration or
EXIT
to
leave the calibration
mode without storing
the values.
14.
Condition the
electrode in deionized
water for 10 minutes.
15.
Remove the
electrode from the
deionized water and
prime the electrode by
clicking the pump until
fresh gel emerges.
FILL
LINE
READ
enter
READ
enter