14
Installation
3.2 Unpack the instrument
CAUTION
Pay attention to the weight (approximately 31 kg) of the
instrument. Do not try to carry the instrument without
assistance. Use only suitable lifting tackle for transport.
Open the shipping container while it is on its end and then slide the
analyzer out of the cardboard.
The items supplied will vary depending on the order.
Standard items supplied for a minimal configuration:
•
PHOSPHAX sc and User Manual
•
Collecting tray
•
Initial set of reagent and cleaning solution
•
Fastening bracket and angle bracket
•
Accessories for tubing and flow-through variant
•
Plug set
3.3 Mechanical installation
Select a suitable place to install the instrument. Plan the
mechanical installation before positioning posts or drilling holes.
Refer to
Make sure that the fastening has sufficient load bearing capacity
(approximately 160 kg). The wall plugs must be selected and
approved to suit the properties of the wall.
Plan cable and tubing routes to avoid sharp bends and tripping
hazards.
When connecting two analyzers (e. g. for measuring two
parameters with a FILTRAX or Ultrafiltration), plan where the
instruments are to be installed and consider the length of the
heated drain tubing (2 m).
3.3.1 Mount the instrument
The PHOSPHAX sc can be mounted in three different ways:
•
Wall Mount (
),
•
Rail Mount; refer to the instruction sheet supplied with the
Rail Mounting Hardware.
•
Stand Mount; refer to the instruction sheet supplied with the
Stand Mounting Hardware.
3.3.1.1 Wall mount
Refer to
, and following instructions to mount the
analyzer to a wall.
1.
Align and install the fastening bracket to the wall.
2.
Attach the angle bracket to the instrument using the
supplied screws.
3.
Slide the bottom of the enclosure onto the fastening bracket.
Содержание PHOSPHAX indoor sc
Страница 2: ......
Страница 6: ...6 Specifications Figure 1 Instrument dimensions PHOSPHAX sc...
Страница 8: ...8 Specifications...
Страница 18: ...18 Installation Figure 8 Open the PHOSPHAX indoor sc enclosure 1 Latches 3 Lock with key 2 Door hook 1 3 2...
Страница 34: ...34 System Startup...
Страница 54: ...54 Replacement Parts and Accessories Exploded view drawings Figure 20 Analyzer enclosure PHOSPHAX sc...
Страница 55: ...55 Replacement Parts and Accessories Figure 21 Analyzer enclosure PHOSPHAX indoor sc...
Страница 56: ...56 Replacement Parts and Accessories Figure 22 Analyzer panel overview...
Страница 57: ...57 Replacement Parts and Accessories Figure 23 Analyzer panel front view details...
Страница 58: ...58 Replacement Parts and Accessories Figure 24 Analyzer panel back view details...
Страница 62: ...62 Limited Warranty...
Страница 96: ...96 Plumbing and connection options...