Installation
3.4.2 Remove the shipping transport locks
Prior to system start-up, the shipping transport locks must be
removed from the sc analyzer.
CAUTION
The enclosure may tip forwards if it has not been fixed in
place. Only open the enclosure if the enclosure is properly
mounted.
Important note:
The cuvette insulation for the measuring unit is
NOT a transport lock. Do NOT remove the cover on the measuring
unit.
1.
Open the enclosure door and secure with the door hook.
2.
Remove the transport lock on the analyzer panel (
Note:
If the instrument is operated with a Filter Probe sc, it is equipped
with an internal compressor.
3.
Remove the cable tie and pull out the compressor transport
lock to the left (
).
Note:
Keep the transport locks for transporting and storage.
Figure 9 Remove the analyzer panel transport locks
1
Transport lock
Содержание PHOSPHAX indoor sc
Страница 2: ......
Страница 6: ...6 Specifications Figure 1 Instrument dimensions PHOSPHAX sc...
Страница 8: ...8 Specifications...
Страница 18: ...18 Installation Figure 8 Open the PHOSPHAX indoor sc enclosure 1 Latches 3 Lock with key 2 Door hook 1 3 2...
Страница 34: ...34 System Startup...
Страница 54: ...54 Replacement Parts and Accessories Exploded view drawings Figure 20 Analyzer enclosure PHOSPHAX sc...
Страница 55: ...55 Replacement Parts and Accessories Figure 21 Analyzer enclosure PHOSPHAX indoor sc...
Страница 56: ...56 Replacement Parts and Accessories Figure 22 Analyzer panel overview...
Страница 57: ...57 Replacement Parts and Accessories Figure 23 Analyzer panel front view details...
Страница 58: ...58 Replacement Parts and Accessories Figure 24 Analyzer panel back view details...
Страница 62: ...62 Limited Warranty...
Страница 96: ...96 Plumbing and connection options...