44
Maintenance
5.
Enter the number of measurements which should be used for
the validation measurements.
(Default value: 3; value range: 2 to 10)
6.
Select START after adjusting both parameters and the analyzer
is going into the service state. The remaining time is displayed
in seconds.
The OUTMODE is set to HOLD.
7.
Select ENTER to modify the analyzer (
a.
Unscrew the fitting (item
) of the sample tube (item
)
which connects the overflow vessel (item
) and the valve
) at the overflow vessel.
b.
Screw in the blind plug (item
) in the thread of the overflow
vessel (item
) and insert the sample tube in a beaker (for
example 150 mL) with standard solution for the validation.
Note:
To receive stable measurement values, close the door of the
analyzer.
8.
Press ENTER to start the validation.
Note:
The remaining time is displayed in seconds:
9.
Press ENTER to escape.
The results are displayed to note down.
•
The discharge value and the Conc value are counted down to
zero.
•
The validation is finished when the process shows the service
mode and the remaining time is 0 seconds.
•
For the adjusted number of validation measurements, the
values are listed and the calculated average of this value is
displayed.
Note:
The data log records the validation values and average value from
the analyzer.
10.
Press ENTER to proceed.
Note:
Press START to confirm the inquiry to return to the
measurement process or to the service mode.
11.
Select ENTER and modify the instrument to the original
analyzer configuration.
12.
Start the measurement mode or hold the service mode.
(
Discharge value
+ measurement value )
×
5 minutes =
remaining time
sec
Содержание PHOSPHAX indoor sc
Страница 2: ......
Страница 6: ...6 Specifications Figure 1 Instrument dimensions PHOSPHAX sc...
Страница 8: ...8 Specifications...
Страница 18: ...18 Installation Figure 8 Open the PHOSPHAX indoor sc enclosure 1 Latches 3 Lock with key 2 Door hook 1 3 2...
Страница 34: ...34 System Startup...
Страница 54: ...54 Replacement Parts and Accessories Exploded view drawings Figure 20 Analyzer enclosure PHOSPHAX sc...
Страница 55: ...55 Replacement Parts and Accessories Figure 21 Analyzer enclosure PHOSPHAX indoor sc...
Страница 56: ...56 Replacement Parts and Accessories Figure 22 Analyzer panel overview...
Страница 57: ...57 Replacement Parts and Accessories Figure 23 Analyzer panel front view details...
Страница 58: ...58 Replacement Parts and Accessories Figure 24 Analyzer panel back view details...
Страница 62: ...62 Limited Warranty...
Страница 96: ...96 Plumbing and connection options...