Table 2  Custom buffer set

Buffer set values Option

Description

Std1 
Std2 
Std3 
Std4 
Std5 

1.09pH—25 °C Pre-set temperature compensated buffer values.

Note: Selected standards must differ by a minimum of 2 pH units. For example, if 1.09 pH is
chosen for the first standard, the second standard must differ by at least 2 pH. Standards that do
not meet this minimum will appear gray on the screen and will not be selectable.

1.68pH—25 °C

4.01pH—25 °C

4.65pH—25 °C

6.86pH—25 °C

7.00pH—25 °C

9.18pH—25 °C

9.23pH—25 °C

10.01pH—
25 °C

12.45pH—
25 °C

Custom Buffer

Custom buffer value.
Range = 2.000 pH to 14.000 pH.
Custom buffer values are not temperature compensated. Custom buffers should be read
at 25 °C.

No Buffer

Standard is undefined when this option is selected.

Change check standard options

1.

Make sure a probe is connected to the meter.

2.

Push   and select PHC281 Settings.

3.

Select Modify Current Settings.

4.

Select Check Standards Options and update the settings:

Option

Description

Standard

Sets the temperature compensated buffer value for check standard—

4.01pH—25 °C

4.65pH—25 °C

6.86pH—25 °C

7.00pH—25 °C

9.18pH—25 °C

9.23pH—25 °C

10.01pH—25 °C

12.45pH—25 °C

Custom

The standard value is shown on the Check Standard Options screen.
No temperature compensation for custom buffers.

Standard
Value

When Standard is set to Custom, enter the standard value using the up/down
arrow keys.

5.

Select Check Standard Reminder and update the settings:

Option

Description

Reminder Repeat

Sets the time interval for the check standard reminder—Off, 30 minutes,
2 h, 4 h, 8 h, 12 h or 24 h.

Allow Defer

Allows the postponement of check standard reminders—Yes or No.

9

Содержание PHC28101

Страница 1: ...combination probe with refillable double junction reference and built in temperature sensor pH range pH 0 to 14 pH resolution User selectable stabilization time and resolution Fast 0 1 Fast 0 01 Medi...

Страница 2: ...viscous storage filling solution completely Blot dry with a lint free cloth 4 Remove the protective tape from the filling hole before initial use refer to Figure 1 on page 2 Dispose of the protective...

Страница 3: ...ne with a pH above and one with a pH below the expected sample pH For a one point calibration select the buffer nearest to the expected sample pH The calibration is recorded in the probe and the data...

Страница 4: ...rect service life time stamp Set the date and time in the meter before the probe is attached If complete traceability is necessary enter a sample ID and operator ID before measurement Refer to the HQd...

Страница 5: ...eat steps 2 4 for additional measurements 7 When measurements are done store the probe refer to Storage on page 10 Run a check standard The run check standard feature validates instrument performance...

Страница 6: ...question mark until the probe is recalibrated To correct the probe calibration and status indicator calibrate the probe refer to Calibration on page 3 Advanced operation Parameter specific settings c...

Страница 7: ...elect Measurement Options and update the settings Option Description Resolution Sets the resolution 0 1pH Fast 0 01pH Fast default 0 01pH Medium 0 01pH Slow or 0 001pH Slow The resolution affects the...

Страница 8: ...ration points necessary before a calibration can be completed 1 default 2 or 3 Slope Limit Sets the slope limit 1 to 10 acceptable slope criteria default 5 The slope must fall within set limits for su...

Страница 9: ...ed when this option is selected Change check standard options 1 Make sure a probe is connected to the meter 2 Push and select PHC281 Settings 3 Select Modify Current Settings 4 Select Check Standards...

Страница 10: ...ute in deionized water 4 Soak the probe in pH 4 buffer for up to 20 minutes then rinse with deionized water 5 Blot dry with a lint free cloth For fats grease and oils 1 Soak the glass bulb in a warm d...

Страница 11: ...rer s website Refer to the HQd Series meter manual for specific instructions for the meter model HQd meter does not support IntelliCAL probe Contact a Technical Support Representative Connect a probe...

Страница 12: ...ts by changing the PHC281 calibration settings and method 2 Recalibrate the probe 3 Run a check standard to check the performance of the probe Drifting Inaccurate readings Contaminated glass bulb Clea...

Страница 13: ...the current method Create a new method with an expanded range Temperature out of range Temperature value is outside of range Make sure that the sample temperature is within the range of the probe Mak...

Страница 14: ...hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info hach lange de www hach lange de HACH LANGE S rl 6 route de Compois 1222 V...

Отзывы: