Hach PHC28101 Скачать руководство пользователя страница 8

4.

Select Calibration Options and update the settings:

Option

Description

Buffer Set

Sets the temperature compensated buffer set used for calibration—

Color Coded—4.01, 7.00, 10.01 (default)

IUPAC—4.01, 7.00, 10.01, 12.45 

DIN—4.65, 9.23 

IUPAC—4.01, 6.86, 10.01, 12.45 

IUPAC—4.01, 6.86, 9.18, 12.45 

IUPAC—4.01, 7.00, 9.18, 12.45 

Custom Buffer Set (refer to 

Table 2

)

Custom buffer sets are characterized at 25 °C (77 °F).
Buffer set values are shown on the Calibration Options screen.

Note: Only the minimum calibration points must be measured for DONE to

be shown on the calibration screen.

Buffer Set
Values

If the Buffer Set is set to Customer Buffer Set, sets the custom buffer set
(refer to 

Table 2

).

Minimum Cal
Points

Sets the minimum number of calibration points necessary before a
calibration can be completed—1 (default), 2 or 3.

Slope Limit

Sets the slope limit—1% to 10% (acceptable slope criteria, default = 5%).
The slope must fall within set limits for successful calibration.

5.

Select Calibration Reminder and update the settings:

Option

Description

Reminder
Repeat

Meter will make an audible sound when a calibration is due and repeat the
sound at the selected interval—Off (default), 2 h, 4 h, 8 h, 2 d, 5 d or 7 d.

Expires

Calibration expires after the selected time—Immediately, Re
30 min, Re 1 h, Re 2 h or Continue Reading.

Note: The meter cannot be used to read samples after calibration has

expired unless Continue Reading is selected.

6.

If prompted, enter a name for the new method settings. Additional changes made to
the settings of an existing method are automatically saved with the same method
name.

7.

Push 

EXIT

 until the meter returns to the measurement mode.

8

Содержание PHC28101

Страница 1: ...combination probe with refillable double junction reference and built in temperature sensor pH range pH 0 to 14 pH resolution User selectable stabilization time and resolution Fast 0 1 Fast 0 01 Medi...

Страница 2: ...viscous storage filling solution completely Blot dry with a lint free cloth 4 Remove the protective tape from the filling hole before initial use refer to Figure 1 on page 2 Dispose of the protective...

Страница 3: ...ne with a pH above and one with a pH below the expected sample pH For a one point calibration select the buffer nearest to the expected sample pH The calibration is recorded in the probe and the data...

Страница 4: ...rect service life time stamp Set the date and time in the meter before the probe is attached If complete traceability is necessary enter a sample ID and operator ID before measurement Refer to the HQd...

Страница 5: ...eat steps 2 4 for additional measurements 7 When measurements are done store the probe refer to Storage on page 10 Run a check standard The run check standard feature validates instrument performance...

Страница 6: ...question mark until the probe is recalibrated To correct the probe calibration and status indicator calibrate the probe refer to Calibration on page 3 Advanced operation Parameter specific settings c...

Страница 7: ...elect Measurement Options and update the settings Option Description Resolution Sets the resolution 0 1pH Fast 0 01pH Fast default 0 01pH Medium 0 01pH Slow or 0 001pH Slow The resolution affects the...

Страница 8: ...ration points necessary before a calibration can be completed 1 default 2 or 3 Slope Limit Sets the slope limit 1 to 10 acceptable slope criteria default 5 The slope must fall within set limits for su...

Страница 9: ...ed when this option is selected Change check standard options 1 Make sure a probe is connected to the meter 2 Push and select PHC281 Settings 3 Select Modify Current Settings 4 Select Check Standards...

Страница 10: ...ute in deionized water 4 Soak the probe in pH 4 buffer for up to 20 minutes then rinse with deionized water 5 Blot dry with a lint free cloth For fats grease and oils 1 Soak the glass bulb in a warm d...

Страница 11: ...rer s website Refer to the HQd Series meter manual for specific instructions for the meter model HQd meter does not support IntelliCAL probe Contact a Technical Support Representative Connect a probe...

Страница 12: ...ts by changing the PHC281 calibration settings and method 2 Recalibrate the probe 3 Run a check standard to check the performance of the probe Drifting Inaccurate readings Contaminated glass bulb Clea...

Страница 13: ...the current method Create a new method with an expanded range Temperature out of range Temperature value is outside of range Make sure that the sample temperature is within the range of the probe Mak...

Страница 14: ...hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info hach lange de www hach lange de HACH LANGE S rl 6 route de Compois 1222 V...

Отзывы: