
6.
Select Acceptance Criteria and update the settings:
Option
Description
Acceptance Limits
Sets the tolerance limits for check standard—0.005pH (default) to
1.000pH.
Cal Expires on Failure
Recalibration required if check standard fails—Yes or No.
7.
If prompted, enter a name for the new method settings. Additional changes made to
the settings of an existing method are automatically saved with the same method
name.
8.
Push
EXIT
until the meter returns to the measurement mode.
Maintenance
Clean the probe
Clean the probe when:
•
Drifting/inaccurate readings occur as a result of contamination on the glass sensor or
the probe being left dry for extended periods of time.
•
Slow stabilization time occurs as a result of contamination on the glass sensor.
•
A calibration error occurs as a result of contamination on the glass sensor.
For general contaminants:
1.
Rinse the probe with deionized water and blot dry with a lint-free cloth.
2.
Soak the glass bulb for 12 to 16 hours in Hach Electrode Cleaning Solution.
3.
Rinse or soak the probe for 1 minute in deionized water.
4.
Soak the probe in pH 4 buffer for up to 20 minutes, then rinse with deionized water.
5.
Blot dry with a lint-free cloth.
For fats, grease and oils:
1.
Soak the glass bulb in a warm detergent solution for up to 2 hours.
2.
Rinse or soak the probe for 1 minute in deionized water.
3.
Soak the probe in pH 4 buffer for up to 20 minutes, then rinse with deionized water.
4.
Blot dry with a lint-free cloth.
Fill the probe
To add filling solution to the probe:
1.
If the filling hole is closed, remove the filling-hole cap from the filling hole (refer to
Figure 1
on page 2).
2.
Remove the cap from the tip of the filling solution bottle.
3.
Hold the bottle so that the tip is down. Put the tip of the bottle in the filling hole.
4.
Slowly squeeze the bottle and fill the probe completely.
Note: Fill the probe completely for the best performance.
5.
If the probe will not be used immediately, store the probe (refer to
Storage
on page 10).
6.
Keep the filling solution bottle and cap for later use.
Note: If the dispensing tip becomes clogged, remove the dispensing tip and soak the tip in
warm water. Dry thoroughly and assemble.
Storage
For the best probe performance, do not let the reference junction dry out.
Short-term and long-term storage
10