Hach PHC28101 Скачать руководство пользователя страница 10

6.

Select Acceptance Criteria and update the settings:

Option

Description

Acceptance Limits

Sets the tolerance limits for check standard—0.005pH (default) to
1.000pH.

Cal Expires on Failure

Recalibration required if check standard fails—Yes or No.

7.

If prompted, enter a name for the new method settings. Additional changes made to
the settings of an existing method are automatically saved with the same method
name.

8.

Push 

EXIT

 until the meter returns to the measurement mode.

Maintenance

Clean the probe

Clean the probe when:

Drifting/inaccurate readings occur as a result of contamination on the glass sensor or
the probe being left dry for extended periods of time.

Slow stabilization time occurs as a result of contamination on the glass sensor.

A calibration error occurs as a result of contamination on the glass sensor.

For general contaminants:

1.

Rinse the probe with deionized water and blot dry with a lint-free cloth.

2.

Soak the glass bulb for 12 to 16 hours in Hach Electrode Cleaning Solution.

3.

Rinse or soak the probe for 1 minute in deionized water.

4.

Soak the probe in pH 4 buffer for up to 20 minutes, then rinse with deionized water.

5.

Blot dry with a lint-free cloth.

For fats, grease and oils:

1.

Soak the glass bulb in a warm detergent solution for up to 2 hours.

2.

Rinse or soak the probe for 1 minute in deionized water.

3.

Soak the probe in pH 4 buffer for up to 20 minutes, then rinse with deionized water.

4.

Blot dry with a lint-free cloth.

Fill the probe

To add filling solution to the probe:

1.

If the filling hole is closed, remove the filling-hole cap from the filling hole (refer to 

Figure 1

 on page 2).

2.

Remove the cap from the tip of the filling solution bottle.

3.

Hold the bottle so that the tip is down. Put the tip of the bottle in the filling hole.

4.

Slowly squeeze the bottle and fill the probe completely.

Note: Fill the probe completely for the best performance.

5.

If the probe will not be used immediately, store the probe (refer to 

Storage

on page 10).

6.

Keep the filling solution bottle and cap for later use.

Note: If the dispensing tip becomes clogged, remove the dispensing tip and soak the tip in

warm water. Dry thoroughly and assemble.

Storage

For the best probe performance, do not let the reference junction dry out.

Short-term and long-term storage

10

Содержание PHC28101

Страница 1: ...combination probe with refillable double junction reference and built in temperature sensor pH range pH 0 to 14 pH resolution User selectable stabilization time and resolution Fast 0 1 Fast 0 01 Medi...

Страница 2: ...viscous storage filling solution completely Blot dry with a lint free cloth 4 Remove the protective tape from the filling hole before initial use refer to Figure 1 on page 2 Dispose of the protective...

Страница 3: ...ne with a pH above and one with a pH below the expected sample pH For a one point calibration select the buffer nearest to the expected sample pH The calibration is recorded in the probe and the data...

Страница 4: ...rect service life time stamp Set the date and time in the meter before the probe is attached If complete traceability is necessary enter a sample ID and operator ID before measurement Refer to the HQd...

Страница 5: ...eat steps 2 4 for additional measurements 7 When measurements are done store the probe refer to Storage on page 10 Run a check standard The run check standard feature validates instrument performance...

Страница 6: ...question mark until the probe is recalibrated To correct the probe calibration and status indicator calibrate the probe refer to Calibration on page 3 Advanced operation Parameter specific settings c...

Страница 7: ...elect Measurement Options and update the settings Option Description Resolution Sets the resolution 0 1pH Fast 0 01pH Fast default 0 01pH Medium 0 01pH Slow or 0 001pH Slow The resolution affects the...

Страница 8: ...ration points necessary before a calibration can be completed 1 default 2 or 3 Slope Limit Sets the slope limit 1 to 10 acceptable slope criteria default 5 The slope must fall within set limits for su...

Страница 9: ...ed when this option is selected Change check standard options 1 Make sure a probe is connected to the meter 2 Push and select PHC281 Settings 3 Select Modify Current Settings 4 Select Check Standards...

Страница 10: ...ute in deionized water 4 Soak the probe in pH 4 buffer for up to 20 minutes then rinse with deionized water 5 Blot dry with a lint free cloth For fats grease and oils 1 Soak the glass bulb in a warm d...

Страница 11: ...rer s website Refer to the HQd Series meter manual for specific instructions for the meter model HQd meter does not support IntelliCAL probe Contact a Technical Support Representative Connect a probe...

Страница 12: ...ts by changing the PHC281 calibration settings and method 2 Recalibrate the probe 3 Run a check standard to check the performance of the probe Drifting Inaccurate readings Contaminated glass bulb Clea...

Страница 13: ...the current method Create a new method with an expanded range Temperature out of range Temperature value is outside of range Make sure that the sample temperature is within the range of the probe Mak...

Страница 14: ...hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info hach lange de www hach lange de HACH LANGE S rl 6 route de Compois 1222 V...

Отзывы: