Hach PHC28101 Скачать руководство пользователя страница 7

Change measurement options

Methods are groups of default or user-defined settings relevant to specific applications. If
the meter is set to the default method and the Modify Current Settings option is chosen, a
prompt for a new name is shown after the changes are entered. The settings are saved
with this name to distinguish them from the default method settings, which cannot be
changed. A saved method can be used instead of multiple adjustments to the individual
settings. Changes made to a user-defined method are automatically saved with the
existing name. Multiple methods can be saved for the same probe on each meter.

1.

Make sure a probe is connected to the meter.

2.

Push   and select PHC281 Settings.

3.

Select Modify Current Settings.

4.

Select Units. Select pH (default) or mV.

Note: The mV option can be used to find the probe offset in a pH 7 buffer or to measure the

slope. Both units are shown when the detail display mode is selected.

5.

Select Measurement Options and update the settings:

Option

Description

Resolution

Sets the resolution:

0.1pH—Fast

0.01pH—Fast (default)

0.01pH—Medium

0.01pH—Slow, or

0.001pH—Slow

The resolution affects the number of decimal places and the stabilization
time. Higher resolution measurements take more time to stabilize. Slower
stabilization times provide higher accuracy measurements.

Measurement
Limits

Set the measurement limits—Lower limit (default: 0.00 pH) or Upper limit
(default: 14.00 pH).
The measurement limits can be set to match the acceptable values for the
sample. When the measurement is above the upper limit setting or below
the lower limit setting, the meter shows an "Out of limits" message. This
message is an alert to a potential problem with the process conditions.

6.

If prompted, enter a name for the new method settings. Additional changes made to
the settings of an existing method are automatically saved with the same method
name.

7.

Push 

EXIT

 until the meter returns to the measurement mode.

Change calibration options

1.

Make sure a probe is connected to the meter.

2.

Push   and select PHC281 Settings.

3.

Select Modify Current Settings.

7

Содержание PHC28101

Страница 1: ...combination probe with refillable double junction reference and built in temperature sensor pH range pH 0 to 14 pH resolution User selectable stabilization time and resolution Fast 0 1 Fast 0 01 Medi...

Страница 2: ...viscous storage filling solution completely Blot dry with a lint free cloth 4 Remove the protective tape from the filling hole before initial use refer to Figure 1 on page 2 Dispose of the protective...

Страница 3: ...ne with a pH above and one with a pH below the expected sample pH For a one point calibration select the buffer nearest to the expected sample pH The calibration is recorded in the probe and the data...

Страница 4: ...rect service life time stamp Set the date and time in the meter before the probe is attached If complete traceability is necessary enter a sample ID and operator ID before measurement Refer to the HQd...

Страница 5: ...eat steps 2 4 for additional measurements 7 When measurements are done store the probe refer to Storage on page 10 Run a check standard The run check standard feature validates instrument performance...

Страница 6: ...question mark until the probe is recalibrated To correct the probe calibration and status indicator calibrate the probe refer to Calibration on page 3 Advanced operation Parameter specific settings c...

Страница 7: ...elect Measurement Options and update the settings Option Description Resolution Sets the resolution 0 1pH Fast 0 01pH Fast default 0 01pH Medium 0 01pH Slow or 0 001pH Slow The resolution affects the...

Страница 8: ...ration points necessary before a calibration can be completed 1 default 2 or 3 Slope Limit Sets the slope limit 1 to 10 acceptable slope criteria default 5 The slope must fall within set limits for su...

Страница 9: ...ed when this option is selected Change check standard options 1 Make sure a probe is connected to the meter 2 Push and select PHC281 Settings 3 Select Modify Current Settings 4 Select Check Standards...

Страница 10: ...ute in deionized water 4 Soak the probe in pH 4 buffer for up to 20 minutes then rinse with deionized water 5 Blot dry with a lint free cloth For fats grease and oils 1 Soak the glass bulb in a warm d...

Страница 11: ...rer s website Refer to the HQd Series meter manual for specific instructions for the meter model HQd meter does not support IntelliCAL probe Contact a Technical Support Representative Connect a probe...

Страница 12: ...ts by changing the PHC281 calibration settings and method 2 Recalibrate the probe 3 Run a check standard to check the performance of the probe Drifting Inaccurate readings Contaminated glass bulb Clea...

Страница 13: ...the current method Create a new method with an expanded range Temperature out of range Temperature value is outside of range Make sure that the sample temperature is within the range of the probe Mak...

Страница 14: ...hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info hach lange de www hach lange de HACH LANGE S rl 6 route de Compois 1222 V...

Отзывы: