Delete a group
1.
On the Counter Navigation screen, push
GROUP
.
2.
In the Group Settings Management window, select a group.
3.
Push
DELETE
.
Add a location to a group
1.
On the Counter Navigation screen, push
GROUP
.
2.
In the Group Settings Management window, select a group.
3.
Push
SETTINGS
.
4.
Push
ADD LOCATION
.
5.
Navigate to the needed location and select it, then push
OK
.
Note: Push the UP or DOWN arrow to change the order of locations in the Group Settings screen. Refer to
on page 26 for more information.
Sizes function
N O T I C E
If a selected channel has not been calibrated, particle counts will be based on interpolation. A maximum of ± 15%
size error and a ± 30% counter error can occur when samples are taken at an interpolated size point.
The Sizes function is a separate option that must be enabled by the manufacturer. The icon
will not show on the Counter Navigation screen if this option has not been purchased. Customized
channel sizes can be selected at time of order.
With the Sizes function enabled, the particle counter can be configured for the particle sizes that
follow
*
:
• 0.3 to 10-μm range: 0.3, 0.4, 0.5, 0.7, 0.8, 1.0, 2.0, 3.0, 4.0, 5.0, 7.0 and 10.0
• 0.5 to 25-μm range: 0.5, 0.6, 0.7, 0.8, 1.0, 2.0, 3.0, 5.0, 7.0, 10.0, 15.0 and 25.0
Note: Channel sizes 0.3 µm and 25.0 µm cannot be configured together.
Data management
Store partial data
Data can be collected from sample measurements that have been stopped by a user or by a flow
error.
1.
On the Counter Navigation screen, push
SYSTEM.
2.
Select the
Custom
tab.
3.
Select the
STORE PARTIAL DATA
checkbox.
Manage the data buffer
The data buffer allows the data to be preserved or overwritten. In addition, the size of the buffer can
be changed and data can be cleared from the buffer.
Set the data buffer to rotate data
The default buffer setting is fixed. In this setting, no new data can be added to the buffer when it is
full. When the buffer is set to rotate, new data can be added to a full buffer, while the oldest data
record is deleted.
*
Assumes 8-channel option is enabled
28
English