3.
Select PDF, Comma Separated File, Tab Separated File, XML or PortAll.
Note: The PDF option shows only after the PDF option has been set to on. Refer to
on page 29.
4.
If PDF was selected and the settings on the PDF tab under System Settings should be used,
select the box to enable PDF Page and File Break Rules.
5.
If needed, change the default file name and confirm.
6.
If USB Flash Drive was selected, insert the USB drive into the USB host on the front of the
instrument. If the USB Flash Drive is not inserted, a warning will show on the screen. The warning
will change to a confirmation after the USB Flash Drive is inserted.
7.
Push
EXPORT
. The status bar will show the progress.
Note: To stop the export, push
.
About status values in exported data
The status value represents several elements of unit status. Environmental, count and concentration
alarms can be indicated in the status value.
To determine the alarm conditions present in the report, subtract the largest possible value that
represents a bit from the Status value in the exported data. Refer to
.
Table 4 Sample status bit mask definitions
Bit
Value
Definition
Bit
Value
Definition
0
1
Calibration
16
65536
Channel 3 concentration alarm
1
2
Flow
17
131072
Channel 4 concentration alarm
2
4
Temperature
18
262144
Channel 5 concentration alarm
3
8
Relative humidity
19
524288
Channel 6 concentration alarm
4
16
Air velocity
20
1048576
Channel 7 concentration alarm
5
32
System alarm
21
2097152
Channel 8 concentration alarm
6
64
Count alarm
22
4194304
Channel 9 concentration alarm
7
128
Concentration alarm
23
8388608
Channel 10 concentration alarm
8
256
Channel 1 count alarm
24
16777216
Channel 11 concentration alarm
9
512
Channel 2 count alarm
25
33554432
Channel 12 concentration alarm
10
1024
Channel 3 count alarm
26
67108864
Channel 7 count alarm
11
2048
Channel 4 count alarm
27
134217728
Channel 8 count alarm
12
4096
Channel 5 count alarm
28
268435456
Channel 9 count alarm
13
8192
Channel 6 count alarm
29
536870912
Channel 10 count alarm
14
16384
Channel 1 concentration alarm
30
1073741824
Channel 11 count alarm
15
32768
Channel 2 concentration alarm
31
2147483648
Channel 12 count alarm
38