6.
If prompted, enter a name for the new method settings. Additional changes made to
the settings of an existing method are automatically saved with the same method
name.
7.
Push
EXIT
until the meter returns to the measurement mode.
Change check standard options
1.
Make sure a probe is connected to the meter.
2.
Push and select MTC301 Settings.
3.
Select Modify Current Settings.
4.
Select Check Standards Options and update the settings:
Option
Description
Standard
Sets the check standard—
•
ZoBell’s (221 mV – 25 °C) (default)
•
Light’s (135 mV – 25 °C)
•
Custom
Temperature compensated for ZoBell’s solution.
Standard value for check standard.
Standard value is shown on Check Standard Options screen.
Lights solution is characterized at 25 °C.
Custom standard values and temperature are user-defined.
Standard
Value
When Standard is set to Custom, enter the standard value using the up/down
arrow keys— -1200.0 to 1200.0 mV (default: 221.0 mV).
The value and temperature for custom check standard are user-defined.
5.
Select Check Standard Reminder and update the settings:
Option
Description
Reminder Repeat
Sets the time interval for the check standard reminder—Off (default), 1 d,
7 d or 30 d.
Allow Defer
Allows the postponement of check standard reminders—Yes (default) or
No.
6.
Select Acceptance Criteria and update the settings:
Option
Description
Acceptance Limits
Sets the tolerance limits for check standard—-±1 mV to 25 mV
(default: ±10 mV).
Cal Expires on Failure
Recalibration required if check standard fails—Yes or No (default).
7.
If prompted, enter a name for the new method settings. Additional changes made to
the settings of an existing method are automatically saved with the same method
name.
8.
Push
EXIT
until the meter returns to the measurement mode.
Maintenance
Clean the probe
Clean the probe when:
•
Drifting/inaccurate readings or slow stabilization time occurs as a result of
contamination on the platinum disc or the probe being left dry for extended periods of
time.
•
Measurement values are outside the calibration/measurement range of the probe
even after a calibration is done using freshly prepared standards.