Note: After cleaning is done, condition the platinum electrode in representative sample before use.

For general cleaning (including oils, greases and organics):

1.

Rinse the probe with deionized water and blot dry with a lint-free cloth.

2.

Put the probe sensor and platinum disc in Electrode Cleaning Solution or warm
detergent solution for up to 15 minutes.

Note: The platinum disc can be polished using a soft cloth or cotton swab with detergent

solution.

3.

Rinse the probe sensor and platinum disc with deionized water. Blot dry with a lint-
free cloth.

For inorganic deposits:

1.

Put the platinum disc in a solution of 0.1 M hydrochloric or nitric acid solution for up to
15 minutes.

2.

Rinse the probe sensor and platinum disc with deionized water. Blot dry with a lint-
free cloth.

Fill the probe

Add filling solution to the probe when the filling solution level is low (refer to 

Preparation

for use

 on page 2). Refer to 

Specifications

 on page 1 for the applicable filling solution.

1.

If the filling hole is closed, remove the filling-hole cap from the filling hole (refer to 

Product overview

 on page 1).

2.

Remove the cap from the tip of the filling solution bottle.

3.

Hold the bottle so that the tip is down. Put the tip of the bottle in the filling hole.

4.

Slowly squeeze the bottle and fully fill the probe.

Note: Fully fill the probe for the best performance.

5.

Put the probe into storage if not used immediately (refer to 

Storage

 on page 11).

6.

Keep the filling solution bottle and cap for later use.

Note: If the dispensing tip becomes clogged, remove the dispensing tip and soak the tip in

warm water. Fully dry and assemble the tip.

Storage

For the best probe performance, do not let the reference junction dry out.

Short-term and long-term storage

Note: The probe can be stored in a sample for up to 2 hours if the sample pH is not high.

1.

Put the filling-hole cap in the filling hole (refer to 

Figure 1

 on page 2).

2.

Rinse the probe with deionized water. Dry the probe with a lint-free cloth.

3.

Fill the probe soaker bottle half full with Hach Electrode Storage Solution or 3 M
potassium chloride (KCl) solution.

4.

Loosen the soaker bottle cap and put the soaker bottle on the probe.

5.

Turn the soaker bottle cap clockwise to tighten the soaker bottle cap.

6.

Make sure that the solution in the soaker bottle completely covers the reference
junction holes.

11

Содержание Intellical MTC30101

Страница 1: ...ng temperature range 0 to 80 C 32 to 176 F Storage temperature range 5 to 40 C 41 to 104 F Junction Ceramic Reference potential versus Standard Hydrogen Electrode 207 mV at 25 C Fill solution 3M KCl w...

Страница 2: ...ode with deionized water thoroughly to remove the viscous storage filling solution completely Blot dry with a lint free cloth 4 Remove the protective tape from the filling hole before initial use refe...

Страница 3: ...he temperature range of 0 to 30 C 32 to 86 F Light s solution should be read at 25 C 77 F Custom ORP Redox calibration solution values and temperature are user defined The calibration is recorded in t...

Страница 4: ...er to the HQd meter manual for more information Regular calibration is required for the best measurement accuracy refer to Calibration on page 3 Prepare the probe for use refer to Preparation for use...

Страница 5: ...ation record the sample pH and temperature 6 Repeat steps 3 6 for additional measurements 7 When measurements are done store the probe refer to Storage on page 11 Measurement conversion to SHE referen...

Страница 6: ...ation state After each addition of titrant the ORP Redox electrode develops a potential proportional to the logarithm of the ratio of the activities of the two oxidation states At the inflection or en...

Страница 7: ...normality of the sample Nx in equivalents per liter Nx Vt x Nt Vx where Nt normality of titrant Eq L Vt volume of titrant at end point mL Vx Volume of sample mL Run Check Standard The Run Check Stand...

Страница 8: ...available options and how to change them are given in the screens tables and procedures throughout this section The settings that can be changed are shown in Table 2 Table 2 Parameter specific settin...

Страница 9: ...ush and select MTC301 Settings 3 Select Modify Current Settings 4 Select Calibration Options and update the settings Option Description Standard Sets the calibration standard ZoBell s 221mV 25 C Light...

Страница 10: ...stom check standard are user defined 5 Select Check Standard Reminder and update the settings Option Description Reminder Repeat Sets the time interval for the check standard reminder Off default 1 d...

Страница 11: ...uct overview on page 1 2 Remove the cap from the tip of the filling solution bottle 3 Hold the bottle so that the tip is down Put the tip of the bottle in the filling hole 4 Slowly squeeze the bottle...

Страница 12: ...r manual Large number of methods stored on the probe Continue to let the probe connect Do not disconnect the probe Damaged probe Make sure there is connectivity with another probe or meter to confirm...

Страница 13: ...onse time Platinum electrode not conditioned for reducing type samples 1 Make sure that the platinum disc is clean and smooth 2 Put the platinum disc in Reducing Solution for ORP Electrodes for 3 10 m...

Страница 14: ...e probe downward in the solution sample to remove any air bubbles in the reference junction holes Out of limits Check standard value is outside of limits set in the current method Make sure that the s...

Страница 15: ...15...

Страница 16: ...ch com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info hach lange de www hach lange de HACH LANGE S rl 6 route de Compois 1222 V se...

Отзывы: