Troubleshooting
Diagnostic and test menu
The diagnostic and test menu shows current and historical information
about the chlorine analyzer. Refer to
and test menu, push the
MENU
key and select Sensor Setup, Diag/Test.
Table 1 DIAG/TEST menu
Option
Description
GATEWAY INFO
Shows the firmware version, driver version, serial number
and boot version for the controller and the types of
sensors connected to the controller.
CAL DAYS
Shows the number of days since the sensor was last
calibrated.
CAL HISTORY
Shows a list of the times when the sensor was calibrated.
Push
ENTER
to scroll through the entries and view a
summary of the calibration data.
RST CAL HISTORY Resets the sensor calibration history. Requires passcode.
SIGNALS
Shows the sensor measurement signal value in mV.
SENSOR DAYS
Shows the number of days the sensor has been in
operation.
RST SENSORS
Resets the sensor days and calibration days to default.
Requires passcode.
CALIBRATION
Shows the slope and offset values for chlorine and pH (if
pH sensor is used). Shows the offset value for the
temperature (if pH sensor is used).
Error list
Errors may occur for various reasons. The sensor reading on the
measurement screen flashes. All outputs are held when specified in the
controller menu. To show the sensor errors, push the
MENU
key and
select Sensor Diag, Error List. A list of possible errors is shown below.
Table 2 Error list for the sensor
Error
Description
Resolution
CL CAL
REQD
A chlorine calibration and/or pH
calibration is required.
The chlorine and/or pH
measurement has changed
enough to cause a Cal Watch
alarm to occur. Refer to the
chlorine sensor manual for
more information.
Calibrate the chlorine sensor
and/or pH sensor.
CL MAINT
REQD
Chlorine sensor maintenance is
required.
The slope is less than 30% or
more than 300% of the default.
The default slope is
100 mV/ppm (100%).
Do sensor maintenance and then
repeat the calibration, or replace
the sensor. Refer to the chlorine
sensor manual for more
information.
CL MAINT
REQD
Chlorine sensor maintenance is
required.
The offset is too high (more
than ±50 mV).
Do sensor maintenance and then
repeat the calibration, or replace
the sensor. Refer to the chlorine
sensor manual for more
information.
CONC
TOO LOW
The chlorine signal is > 0 mV.
The potential delivered by the
sensor is outside of the range
0 to -2500 mV.
A connection error occurred or
the chlorine sensor is not
polarized/equilibrated in the
sample sufficiently.
CONC
TOO HIGH
The chlorine signal is less than
-2500 mV or greater than
2500 mV (by absolute value).
Warning list
A warning does not affect the operation of menus, relays and outputs. A
warning icon flashes and a message is shown on the bottom of the
measurement screen. To show the sensor warnings, push the
MENU
20
English
Содержание Chlorine Sensor
Страница 8: ...8 English...
Страница 17: ...English 17...
Страница 29: ...Deutsch 29...
Страница 51: ...Assemblare il sensore per cloro Italiano 51...
Страница 53: ...Italiano 53...
Страница 73: ...Assemblage du capteur de chlore Fran ais 73...
Страница 75: ...Fran ais 75...
Страница 95: ...Montaje del sensor de cloro Espa ol 95...
Страница 97: ...Espa ol 97...
Страница 105: ...Espa ol 105...
Страница 117: ...Montar o sensor de cloro Portugu s 117...
Страница 119: ...Portugu s 119...
Страница 139: ...e tina 139...
Страница 160: ...160 Dansk...
Страница 181: ...Chloorsensor samenbouwen Nederlands 181...
Страница 183: ...Nederlands 183...
Страница 203: ...Monta czujnika chloru Polski 203...
Страница 205: ...Polski 205...
Страница 225: ...Svenska 225...
Страница 235: ...Svenska 235...
Страница 242: ...12 VDC 30 mA 5 45 C 41 113 F 0 50 C 32 122 F 15 25 C 59 77 F 195 7 68 25 0 98 1 5 M12 1 2 PVC 1 2 242...
Страница 243: ...ESD 10 pH CLF10sc CLT10sc sc 1 243...
Страница 244: ...1 1 4 2 5 3 6 30 mV 244...
Страница 245: ...pH 2 2 1 4 2 5 3 O MSDS 48 245...
Страница 246: ...48 246...
Страница 247: ...247...
Страница 248: ...6 12 260 248...
Страница 249: ...0 1 ppm mg L Configure 1 MENU Sensor Setup Calibrate Cal Options EDIT NAME 10 249...
Страница 257: ...6 1 3 6 3 1 1 6 257...
Страница 258: ...1 2 3 245 4 M48 5 6 7 3 8 9 10 6 12 260 11 258...
Страница 259: ...3 1 3 2 4 1 2 3 245 4 5 6 7 8 6 12 260 9 1 2 3 245 4 M48 5 6 1 2 7 8 9 260 10 259...
Страница 270: ...270 Magyar...
Страница 280: ...280 Magyar...
Страница 291: ...Asamblarea senzorului de clor Rom n 291...
Страница 293: ...Rom n 293...
Страница 309: ...12 30 5 45 C 41 113 F 0 50 C 32 122 F 15 25 C 59 77 F 195 7 68 25 0 98 1 5 M12 1 2 1 2 309...
Страница 310: ...pH 10 pH CLF10sc CLT10sc sc 1 1 1 4 2 5 3 6 310...
Страница 311: ...OFF 30 pH pH 2 311...
Страница 312: ...2 1 4 2 5 3 MSDS M48 M48 312...
Страница 313: ...313...
Страница 314: ...6 12 327 314...
Страница 315: ...315...
Страница 324: ...1 1 324...
Страница 325: ...1 2 3 312 4 M48 5 6 7 3 8 9 10 6 12 327 11 325...
Страница 326: ...3 1 3 2 4 1 2 3 312 4 5 6 7 8 6 12 327 9 1 2 3 312 4 M48 5 6 1 2 7 8 9 327 10 326...
Страница 327: ...MSDS 3 308 1 2 3 4 1 312 2 3 4 5 6 7 6 12 327 8 1 1 GATEWAY INFO RST CAL HISTORY SIGNALS 327...
Страница 328: ...1 RST SENSORS CALIBRATION pH pH pH 2 CL CAL REQD CL pH pH Cal Watch pH CL MAINT REQD CL 30 300 100 100 328...
Страница 337: ...T rk e 337...
Страница 358: ...358 Slovensk jazyk...
Страница 368: ...368 Slovensk jazyk...
Страница 379: ...Slovenski 379...
Страница 400: ...400 Hrvatski...
Страница 417: ...1 2 PVC 1 2 417...
Страница 418: ...pH 10 pH CLF10sc CLT10sc sc 1 1 1 4 2 5 3 6 LED LED LED 418...
Страница 419: ...LED LED 30 mV pH pH 2 2 1 4 2 5 3 O ring 419...
Страница 420: ...MSDS M48 M48 O ring 420...
Страница 421: ...421...
Страница 422: ...6 12 435 422...
Страница 423: ...423...
Страница 432: ...1 1 432...
Страница 433: ...1 2 3 420 4 M48 5 6 7 3 8 9 10 6 12 435 11 433...
Страница 434: ...3 1 3 2 4 1 2 3 420 4 5 6 7 8 6 12 435 9 1 2 3 420 4 M48 5 6 1 2 7 8 9 435 10 434...
Страница 440: ......
Страница 441: ......