4.
Push the
ENTER
key and select Activate TMR.
5.
Set the amount of time a measurement(s) can be outside the
deviation range before an alarm occurs:
Option
Description
ALL
ACTIVATE TMR: Sets the amount of time the chlorine AND pH
measurements can be out of range before an alarm occurs-10 to
99 minutes (10 minutes default).
CONFID LEVEL: Set the percentage rate at which chlorine AND
pH measurements must be out of range before an alarm occurs
and must be in range before an alarm is automatically canceled
by the instrument-50 to 95% (default).
CL2/pH
ONLY
ACTIVATE TMR: Sets the amount of time the chlorine OR pH
measurements can be out of range before an alarm occurs-10 to
999 minutes (30 minutes default).
CONFID LEVEL: Set the percentage rate at which chlorine OR
pH measurements must be out of range before an alarm occurs
and must be in range before an alarm is automatically canceled
by the instrument-50 to 95% (default).
LCC
ACTIVATE TMR: Set the amount of time a chlorine measurement
can be 0.5 ppm or higher before an alarm occurs-10 to
999 minutes (30 minutes default). An alarm only occurs if the
previous chlorine sensor calibration was done using a process
flow with a low chlorine concentration (LCC) (< 0.5 ppm).
CONFID LEVEL: Set the percentage rate at which chlorine
measurements must be 0.5 ppm or higher before an alarm
occurs and must be less than 0.5 ppm before an alarm is
automatically canceled by the instrument-50 to 95% (default).
6.
Push the
ENTER
key and select Deactivate TMR.
7.
Set the amount of time an alarm is on before it is automatically
canceled (deactivated) by the instrument if the measurements go
back in to range at the percentage rate (confidence level) selected in
Activate TMR:
Option
Description
ALL
Sets the amount of time a chlorine AND pH measurement
deviation alarm is on before it is canceled-10 to 99 minutes
(30 minutes default).
Option
Description
CL2/pH
ONLY
Sets the amount of time a chlorine OR pH measurement
deviation alarm is on before it is canceled-10 to 999 minutes
(30 minutes default).
LCC
Sets the amount of time an LCC alarm is on before it is
canceled-10 to 999 minutes (30 minutes default).
8.
Push the
ENTER
key and select CL2 Deviation.
9.
Set the chlorine measurement deviation values that activate alarms:
Option
Description
CL2 ERR DEV
Set the chlorine measurement deviation that will activate an
error alarm-30 to 99% (50% default).
CL2 WRN DEV
Set the chlorine measurement deviation that will activate a
warning alarm-10 to 30% (20% default).
Note: Chlorine deviation is measured by the instrument using the chlorine
value recorded during the last process concentration calibration for chlorine.
10.
Push the
ENTER
key and select pH Deviation.
11.
Set the pH measurement deviation values that activate alarms:
Option
Description
pH ERR DEV
Set the pH measurement deviation that will activate an error
alarm-1 (default) to 3 pH units.
pH WRN DEV
Set the pH measurement deviation that will activate a
warning alarm-0.5 (default) to 1 pH unit.
Note: pH deviation is measured by the instrument using the pH value recorded
during the last process concentration calibration for chlorine.
Calibrate the sensor
About sensor calibration
The sensor characteristics slowly shift over time and cause the sensor to
lose accuracy. The sensor must be calibrated regularly to maintain
accuracy. The calibration frequency varies with the application and is
best determined by experience.
12
English
Содержание Chlorine Sensor
Страница 8: ...8 English...
Страница 17: ...English 17...
Страница 29: ...Deutsch 29...
Страница 51: ...Assemblare il sensore per cloro Italiano 51...
Страница 53: ...Italiano 53...
Страница 73: ...Assemblage du capteur de chlore Fran ais 73...
Страница 75: ...Fran ais 75...
Страница 95: ...Montaje del sensor de cloro Espa ol 95...
Страница 97: ...Espa ol 97...
Страница 105: ...Espa ol 105...
Страница 117: ...Montar o sensor de cloro Portugu s 117...
Страница 119: ...Portugu s 119...
Страница 139: ...e tina 139...
Страница 160: ...160 Dansk...
Страница 181: ...Chloorsensor samenbouwen Nederlands 181...
Страница 183: ...Nederlands 183...
Страница 203: ...Monta czujnika chloru Polski 203...
Страница 205: ...Polski 205...
Страница 225: ...Svenska 225...
Страница 235: ...Svenska 235...
Страница 242: ...12 VDC 30 mA 5 45 C 41 113 F 0 50 C 32 122 F 15 25 C 59 77 F 195 7 68 25 0 98 1 5 M12 1 2 PVC 1 2 242...
Страница 243: ...ESD 10 pH CLF10sc CLT10sc sc 1 243...
Страница 244: ...1 1 4 2 5 3 6 30 mV 244...
Страница 245: ...pH 2 2 1 4 2 5 3 O MSDS 48 245...
Страница 246: ...48 246...
Страница 247: ...247...
Страница 248: ...6 12 260 248...
Страница 249: ...0 1 ppm mg L Configure 1 MENU Sensor Setup Calibrate Cal Options EDIT NAME 10 249...
Страница 257: ...6 1 3 6 3 1 1 6 257...
Страница 258: ...1 2 3 245 4 M48 5 6 7 3 8 9 10 6 12 260 11 258...
Страница 259: ...3 1 3 2 4 1 2 3 245 4 5 6 7 8 6 12 260 9 1 2 3 245 4 M48 5 6 1 2 7 8 9 260 10 259...
Страница 270: ...270 Magyar...
Страница 280: ...280 Magyar...
Страница 291: ...Asamblarea senzorului de clor Rom n 291...
Страница 293: ...Rom n 293...
Страница 309: ...12 30 5 45 C 41 113 F 0 50 C 32 122 F 15 25 C 59 77 F 195 7 68 25 0 98 1 5 M12 1 2 1 2 309...
Страница 310: ...pH 10 pH CLF10sc CLT10sc sc 1 1 1 4 2 5 3 6 310...
Страница 311: ...OFF 30 pH pH 2 311...
Страница 312: ...2 1 4 2 5 3 MSDS M48 M48 312...
Страница 313: ...313...
Страница 314: ...6 12 327 314...
Страница 315: ...315...
Страница 324: ...1 1 324...
Страница 325: ...1 2 3 312 4 M48 5 6 7 3 8 9 10 6 12 327 11 325...
Страница 326: ...3 1 3 2 4 1 2 3 312 4 5 6 7 8 6 12 327 9 1 2 3 312 4 M48 5 6 1 2 7 8 9 327 10 326...
Страница 327: ...MSDS 3 308 1 2 3 4 1 312 2 3 4 5 6 7 6 12 327 8 1 1 GATEWAY INFO RST CAL HISTORY SIGNALS 327...
Страница 328: ...1 RST SENSORS CALIBRATION pH pH pH 2 CL CAL REQD CL pH pH Cal Watch pH CL MAINT REQD CL 30 300 100 100 328...
Страница 337: ...T rk e 337...
Страница 358: ...358 Slovensk jazyk...
Страница 368: ...368 Slovensk jazyk...
Страница 379: ...Slovenski 379...
Страница 400: ...400 Hrvatski...
Страница 417: ...1 2 PVC 1 2 417...
Страница 418: ...pH 10 pH CLF10sc CLT10sc sc 1 1 1 4 2 5 3 6 LED LED LED 418...
Страница 419: ...LED LED 30 mV pH pH 2 2 1 4 2 5 3 O ring 419...
Страница 420: ...MSDS M48 M48 O ring 420...
Страница 421: ...421...
Страница 422: ...6 12 435 422...
Страница 423: ...423...
Страница 432: ...1 1 432...
Страница 433: ...1 2 3 420 4 M48 5 6 7 3 8 9 10 6 12 435 11 433...
Страница 434: ...3 1 3 2 4 1 2 3 420 4 5 6 7 8 6 12 435 9 1 2 3 420 4 M48 5 6 1 2 7 8 9 435 10 434...
Страница 440: ......
Страница 441: ......