6
English
Operation
Successive
measurements channel
1
Number of measurements in succession ( = measurements on channel 1 +
DISCHARGE VAL 1 channel 1). Available with the 2-channel version.
Discarded
measurements channel
1
Number of discarded values after switching from channel 1 to channel 2. Available with the
2-channel version.
Location 2
Settings for location 2
Name
Enter the name for the measuring location as required. Available with the 2-channel version.
Parameter
Select output: ammonium or ammonium nitrogen. Available with the 2-channel version.
Unit
Select output: mg/L or ppm. Available with the 2-channel version.
Successive
measurements channel
2
Number of measurements in succession ( = measurements on channel 2 +
DISCHARGE VAL 2 channel 2). Available with the 2-channel version.
Discarded
measurements channel
2
Number of discarded values after switching from channel 2 to channel 1. Available with the
2-channel version.
Measurement schedule
Interval
Enter measurement interval (Time interval between two measurements). ATTENTION with
filtration probe and 5 min operation: increased pump speed in the filtration probe, annual
filtration probe maintenance is necessary.
Start by bus
Start by bus
YES/NO; option whether the instrument is measuring continuously or measurements are
triggered by field-bus. "Fieldbus" must be activated under "TEST/MAINT". Instrument will
switch to 5 min interval when activated.
Measurements
Number of measurements after a bus activation.
Discarded
measurements
Number of discharged values that precede the measurements.
Average
Number of measurements that are averaged. (Affects only the bus triggered measurements)
Cleaning
Interval
Number of hours between the cleanings
Start
Start time for the cleaning (In case of more than one cleaning per day: Enter start time for the
first cleaning.)
Discarded
measurements
Number of measured values that are discarded after a cleaning operation.
Cleaning output mode
Value that is output during a cleaning operation and the following discarded values.
HOLD = last measured value, SET TRANSFER = value to be entered.
Cuvette temperature
Temperature for cuvette and electrode
Recommended: Select 45 °C at air temperature: up to 35 °C, 50 °C at air temperature: up to
40 °C, 55 °C at air temperature: up to 45 °C, at 55 °C reduced accuracy and shortened
electrode service life may occur.
Tube heating
Heating is on.
Probe tube heating switches on at the start of the selected month when using the Filter
Probe sc.
Heating is off.
Probe tube heating switches off at the end of the selected month when using the Filter
Probe sc.
Reagent warning
Reagent warning
On/Off
When On is selected: determines the warning output if the reagent levels are low
Warning
Determines the level below which the reagent must drop to trigger the warning.
Module status: Warning
Settings (continued)
5.2 Sensor setup menu (continued)
Содержание AMTAX sc
Страница 2: ......
Страница 4: ...4 Table of contents...
Страница 12: ...12 English Operation...
Страница 24: ...24 English Maintenance...
Страница 32: ...32 English Troubleshooting...
Страница 40: ...40 Deutsch Funktion...
Страница 52: ...52 Deutsch Wartung...
Страница 68: ...68 Espa ol Operaci n...
Страница 80: ...80 Espa ol Mantenimiento...
Страница 108: ...108 Fran ais Maintenance...
Страница 124: ...124 Italiano Procedure...
Страница 136: ...136 Italiano Manutenzione...
Страница 152: ...152 Nederlands Bedrijf...
Страница 182: ...182 Dansk Vedligeholdelse 630 35 6 265 3 570 720 35 6 265 1 1700 Table 55 Reng ringsinterval...
Страница 200: ...200 Dansk Fejls gning...
Страница 208: ...208 Polski U ytkowanie...
Страница 218: ...218 Polski Rozwi zywanie Problem w...
Страница 231: ......