26
English
Troubleshooting
Analyzer is too cold.
Instrument goes into
the Maintenance mode
Instrument interior has
been below 4 °C
(39 °F) for more than
5 minutes
Close instrument,
check heating
Reset error manually
TEST/MAINTENANCE
>RESET ERROR
The analyzer is not
heating up.
Instrument goes into
the Maintenance mode
Instrument cannot heat
interior adequately.
(internal temperature
< 20 °C (68 °F) for
30 min)
Close instrument,
check heating
Reset error manually
Cooling failed!
Maintenance mode,
starts automatically
after cooling down
Instrument interior is
too warm (> 47 °C
(117 °F) with electrode
temp. 45 °C (113 °F)
> 52 °C (126 °F) with
electrode temp. 50 °C
(122 °F) or >57 °C
(135 °F) with electrode
temp. 55 °C (130 °F) )
Check air filter and
clean/replace,
check fan.
Manual reset or when
temperature falls 2 °C
(3.6 °F) below limit
Humidity in analyzer
Maintenance mode
There is liquid in the
collecting tray
Identify cause and
rectify
Reset error manually
Humidity in sensor
Maintenance mode,
the filtration probe sc is
isolated from the mains
There is liquid in the
filtration probe
enclosure
Immediately take
filtration probe sc out
of operation and
contact service. Take
the filtration probe sc
out of the tank and
store the filter modules
so that they remain
moist. (See operating
instructions for the
filtration probe sc).
Reset error manually
Filtration probe is
missing.
Maintenance mode,
the filtration probe sc is
isolated from the mains
The filtration probe sc
is faulty or not
connected
Immediately take
filtration probe sc out
of operation and
contact service. Take
the filtration probe sc
out of the tank and
store the filter modules
so that they remain
moist. (See operating
instructions for the
filtration probe sc).
Reset error manually
Temperature sensor is
defective.
Maintenance mode,
fan running, heating off
The temperature
sensor for the internal
instrument temperature
is faulty
Immediately switch off
instrument, contact
service, replace main
circuit board
Reset error manually
Cuvette sensor is
defective.
Maintenance mode,
cuvette heating off
The temperature
sensor for the cuvette
is faulty
Contact service,
replace cuvette/sensor
Reset error manually
Cuvette heating is
defective.
Continued
measurement
The cuvette is not
being heated
adequately
Close instrument door
and wait 10 minutes if
error accurs again,
contact service
Reset error manually
7.2.2 Error messages (continued)
Error displayed
Instrument reaction
Cause
Solution
Reset error
Содержание AMTAX sc
Страница 2: ......
Страница 4: ...4 Table of contents...
Страница 12: ...12 English Operation...
Страница 24: ...24 English Maintenance...
Страница 32: ...32 English Troubleshooting...
Страница 40: ...40 Deutsch Funktion...
Страница 52: ...52 Deutsch Wartung...
Страница 68: ...68 Espa ol Operaci n...
Страница 80: ...80 Espa ol Mantenimiento...
Страница 108: ...108 Fran ais Maintenance...
Страница 124: ...124 Italiano Procedure...
Страница 136: ...136 Italiano Manutenzione...
Страница 152: ...152 Nederlands Bedrijf...
Страница 182: ...182 Dansk Vedligeholdelse 630 35 6 265 3 570 720 35 6 265 1 1700 Table 55 Reng ringsinterval...
Страница 200: ...200 Dansk Fejls gning...
Страница 208: ...208 Polski U ytkowanie...
Страница 218: ...218 Polski Rozwi zywanie Problem w...
Страница 231: ......