5
English
Section 5
Operation
The AMTAX sc can only be operated with an sc1500 controller. For
further information, see the sc1500 User Manual.
An LED on the door indicates the current operating state.
Instrument, chemicals and electrode are temperature-sensitive. To
prevent incorrect measurements, only operate the instrument with
the door closed.
5.1 Sensor diagnostics menu
SELECT AMTAX sc (if more than one sensor or analyzer is attached)
AMTAX sc
Error list
Displays all errors currently present in the sensor
Warning list
Displays all warnings currently present in the sensor
5.2 Sensor setup menu
SELECT AMTAX sc (if more than one sensor or analyzer is attached)
CALIBRATION
Correction factor
Displays the correction factor
Location 1
Displays location1 which is adjusted in the CONFIGURE menu
Gain correction
Adjusts the correction factor for channel 1
Location2
Available with the 2-channel version.
Gain correction
Available with the 2-channel version.
Calibration
Triggers an automatic calibration, then measuring mode
Calibration & cleaning
Triggers an automatic calibration with subsequent automatic cleaning, then measuring mode.
Settings
Calibration output mode
Value that is output during a calibration and the following discarded values.
Hold = last measured value, Transfer = value to be entered.
Automatic calibration
Interval
Time interval between two calibrations
Start
Start time for the calibration (In case of more than one calibration per day: Enter start time for
the first calibration.)
Discarded
measurements
Number of measured values that are discarded after a calibration.
Adjustment method
Select the calibration method
Advanced
Default calibration method for software version
≥
1.60 (For best accuracy with lowest
measuring values.)
Basic
Calibration method for software version
<
1.60 is available for compatiblility reasons (Not
selectable in the measuring range 0.02–5 mg/L.).
Reset
Resets to factory default settings.
Settings
Location 1
Settings for location 1
Name
Enter the name for the measuring location as required.
Parameter
Select output: ammonium or ammonium nitrogen
Unit
Select output: mg/L or ppm
Settings (continued)
Содержание AMTAX sc
Страница 2: ......
Страница 4: ...4 Table of contents...
Страница 12: ...12 English Operation...
Страница 24: ...24 English Maintenance...
Страница 32: ...32 English Troubleshooting...
Страница 40: ...40 Deutsch Funktion...
Страница 52: ...52 Deutsch Wartung...
Страница 68: ...68 Espa ol Operaci n...
Страница 80: ...80 Espa ol Mantenimiento...
Страница 108: ...108 Fran ais Maintenance...
Страница 124: ...124 Italiano Procedure...
Страница 136: ...136 Italiano Manutenzione...
Страница 152: ...152 Nederlands Bedrijf...
Страница 182: ...182 Dansk Vedligeholdelse 630 35 6 265 3 570 720 35 6 265 1 1700 Table 55 Reng ringsinterval...
Страница 200: ...200 Dansk Fejls gning...
Страница 208: ...208 Polski U ytkowanie...
Страница 218: ...218 Polski Rozwi zywanie Problem w...
Страница 231: ......