94
Plumbing and Connection Options
A.12 Option 8a plumbing and connections
Option 8a uses an sc analyzer with the FILTRAX. The waste of the
analyzer is discharged back into an open drain.
Refer to
and the following instructions for Option 8a:
1.
Install the FILTRAX into the sample stream. Refer to the
FILTRAX User Manual for more information.
2.
Feed the heated hose from the FILTRAX through the analyzer
opening (
, item 3). Use Seal Plug #1 to secure.
3.
Feed the drain tube through the analyzer opening (item 5).
Use Seal Plug #3 to secure.
Note:
Tubes can be pushed through prepared holes on Seal Plug #3.
4.
Connect the drain tube to the T-fitting.
5.
Connect the FILTRAX sample line to the bottom inlet on the
overflow vessel using the fittings (item 4).
6.
Feed the drain tube to a lower drain (maximum 2 m/6.5 ft).
Содержание AMTAX sc
Страница 2: ......
Страница 7: ...7 Specifications Figure 1 Instrument dimensions AMTAX sc...
Страница 12: ...12 General Information...
Страница 18: ...18 Installation Figure 8 Open the enclosure AMTAX indoor sc 1 Latches 3 Lock with key 2 Door hook...
Страница 36: ...36 System Start Up...
Страница 44: ...44 Operation...
Страница 56: ...56 Maintenance...
Страница 64: ...64 Troubleshooting...
Страница 68: ...68 Replacement Parts and Accessorie Exploded view drawings Figure 22 Analyzer enclosure AMTAX sc...
Страница 69: ...69 Replacement Parts and Accessorie Figure 23 Analyzer enclosure AMTAX indoor sc...
Страница 70: ...70 Replacement Parts and Accessorie Figure 24 Analyzer panel overview...
Страница 71: ...71 Replacement Parts and Accessorie Figure 25 Analyzer panel front view details...
Страница 72: ...72 Replacement Parts and Accessorie Figure 26 Analyzer panel back view details...
Страница 76: ...76 Warranty and liability...
Страница 110: ...110 Plumbing and Connection Options...