Troubleshooting
ANALYZ. TO COLD
Instrument goes into
the service state
Instrument interior has
been below 4 °C
(39 °F) for more than
5 minutes
Close instrument,
check heating
Reset error manually
TEST/MAINT>RESET
ERROR
NO HEAT UP
Instrument goes into
the service state
Instrument cannot heat
interior adequately.
(internal temperature
< 20 °C (68 °F) for
30 min)
Close instrument,
check heating
Reset error manually
COOLING FAILED
Service state, starts
automatically after
cooling down
Instrument interior is
too warm (> 47 °C
(117 °F) with electrode
temp. 45 °C (113 °F)
> 52 °C (126 °F) with
electrode temp. 50 °C
(122 °F) or >57 °C
(135 °F) with electrode
temp. 55 °C (130 °F) )
Check air filter and
clean/replace,
check fan.
Manual reset or when
temperature falls 2 °C
(3.6 °F) below limit
HUMIDITY ANALY
Service state
There is liquid in the
collecting tray
Identify cause and
rectify
Reset error manually
HUMIDITY PROBE
Service state, the
filtration probe sc is
isolated from the mains
There is liquid in the
filtration probe
enclosure
Immediately take
filtration probe sc out of
operation and contact
service. Take the
filtration probe sc out of
the tank and store the
filter modules so that
they remain moist.
(See operating
instructions for the
filtration probe sc).
Reset error manually
PROBE MISSING
Service state, the
filtration probe sc is
isolated from the mains
The filtration probe sc
is faulty or not
connected
Immediately take
filtration probe sc out of
operation and contact
service. Take the
filtration probe sc out of
the tank and store the
filter modules so that
they remain moist.
(See operating
instructions for the
filtration probe sc).
Reset error manually
TEMPSENS DEFECT
Service state, fan
running, heating off
The temperature
sensor for the internal
instrument temperature
is faulty
Immediately switch off
instrument, contact
service, replace main
circuit board
Reset error manually
CUVSENSOR
DEFECT
Service state, cuvette
heating off
The temperature
sensor for the cuvette
is faulty
Contact service,
replace cuvette/sensor
Reset error manually
CUVHEAT DEFECT
Continued
measurement
The cuvette is not
being heated
adequately
Close instrument door
and wait 10 minutes if
error accurs again,
contact service
Reset error manually
7.2.2 Error messages (continued)
Error displayed
Instrument reaction
Cause
Solution
Reset error
Содержание AMTAX sc
Страница 2: ......
Страница 7: ...7 Specifications Figure 1 Instrument dimensions AMTAX sc...
Страница 12: ...12 General Information...
Страница 18: ...18 Installation Figure 8 Open the enclosure AMTAX indoor sc 1 Latches 3 Lock with key 2 Door hook...
Страница 36: ...36 System Start Up...
Страница 44: ...44 Operation...
Страница 56: ...56 Maintenance...
Страница 64: ...64 Troubleshooting...
Страница 68: ...68 Replacement Parts and Accessorie Exploded view drawings Figure 22 Analyzer enclosure AMTAX sc...
Страница 69: ...69 Replacement Parts and Accessorie Figure 23 Analyzer enclosure AMTAX indoor sc...
Страница 70: ...70 Replacement Parts and Accessorie Figure 24 Analyzer panel overview...
Страница 71: ...71 Replacement Parts and Accessorie Figure 25 Analyzer panel front view details...
Страница 72: ...72 Replacement Parts and Accessorie Figure 26 Analyzer panel back view details...
Страница 76: ...76 Warranty and liability...
Страница 110: ...110 Plumbing and Connection Options...