52
Maintenance
6.6 Validation (Analytical quality assurance)
Regular validation checks of the complete instrument must be
completed to make sure the analysis results are reliable.
Required parts:
•
Blind plug LZY193 (Plugging set LZY007)
•
Beaker (for example 150 mL)
•
Standard solution for validation
Follow the internal menu steps for the validation.
1.
From the MENU select SENSOR SETUP>AMTAX SC and
press ENTER.
2.
Select MAINTENANCE>TEST/MAINT>VALIDATION>
DISCHARGE.
3.
Enter the number of measurements which should be
discharged before starting the measurements of the validation.
(Default value: 3; value range: 2 to 5)
4.
Select NUMBER OF MEAS..
5.
Enter the number of measurements which should be used for
the validation measurements.
(Default value: 3; value range: 2 to 10)
6.
Select START after adjusting both parameters and the analyzer
is going into the service state. The remaining time is displayed
in seconds.
The OUTMODE is set to HOLD.
7.
Select ENTER to modify the analyzer (
a.
Unscrew the fitting (item
) of the sample tube (item
)
which connects the overflow vessel (item
) and the valve
) at the overflow vessel.
b.
Screw in the blind plug (item
) in the thread of the overflow
vessel (item
) and insert the sample tube in a beaker (for
example 150 mL) with standard solution for the validation.
Note:
To receive stable measurement values, close the door of the
analyzer.
8.
Press ENTER to start the validation.
Note:
The remaining time is displayed in seconds:
9.
Press ENTER to escape.
The results are displayed to note down.
•
The discharge value and the Conc value are counted down
to zero.
(Discharge value + measurement value) 5 minutes = remaining time/sec
×
Содержание AMTAX sc
Страница 2: ......
Страница 7: ...7 Specifications Figure 1 Instrument dimensions AMTAX sc...
Страница 12: ...12 General Information...
Страница 18: ...18 Installation Figure 8 Open the enclosure AMTAX indoor sc 1 Latches 3 Lock with key 2 Door hook...
Страница 36: ...36 System Start Up...
Страница 44: ...44 Operation...
Страница 56: ...56 Maintenance...
Страница 64: ...64 Troubleshooting...
Страница 68: ...68 Replacement Parts and Accessorie Exploded view drawings Figure 22 Analyzer enclosure AMTAX sc...
Страница 69: ...69 Replacement Parts and Accessorie Figure 23 Analyzer enclosure AMTAX indoor sc...
Страница 70: ...70 Replacement Parts and Accessorie Figure 24 Analyzer panel overview...
Страница 71: ...71 Replacement Parts and Accessorie Figure 25 Analyzer panel front view details...
Страница 72: ...72 Replacement Parts and Accessorie Figure 26 Analyzer panel back view details...
Страница 76: ...76 Warranty and liability...
Страница 110: ...110 Plumbing and Connection Options...