Specifications
Specifications are subject to change without notice.
General
Controller Dimensions
14.4 cm x 14.4 cm x 11.9 cm (5.6 in. x 5.6 in. x 4.6 in.) (W x H x D)
Mounting
Pre-mounted on panel
Instrument Shipping Weight
4.1 kg (9 lb)
Sample Requirements
Sample Flow Rate to Analyzer
166 to 250 mL/min (10 to 15 L/h) recommended
Inlet Pressure to Instrument
0.5–6 bar (7–89 psi)
Sample Temperature Range
5–45 °C (41–113 °F)
Sample Inlet Tubing: at instrument ¼ in. OD
Drain Fitting
5
/
16
in. OD
Application Sample
Industrial and boiler waters
Electrical
Power Requirements
100–240 V ac ±10%; 90 VA, 50/60 Hz, 2 Amp fuse
Power Connection
Connection made to a three-position terminal block through one of four ½-inch conduit holes
in the case. Wire size: 12–18 AWG.
Installation Category
II
Alarm Contact Rating
Four relays (SPST) each rated at 3 A (resistive), 240 V ac maximum. The contacts are
unpowered.
Use either high voltage (greater than 30 V RMS and 42.2 V PEAK or 60 V dc) or low voltage
(less than 30 V RMS and 42.2 V PEAK or 60 V dc). Do not use a combination of high and low
voltage.
Alarm Connection
Connection made by four removable two-position terminals through one of four
½-inch conduit holes in the case. Wire size: 12–18 AWG.
Analog Output
Two isolated outputs, 0/4–20 mA. Maximum impedance 800 ohm.
Recorder Output Connections
Connection made to a terminal block through one of four ½-inch conduit holes in the case.
Wire size: 12–18 AWG.
Digital Output
RS485
Performance
Measurement Range
0–500 ppb dissolved N
2
H
4
, 0–100 ppb dissolved carbohydrazide
Repeatability
± 2% of measurement or 1 ppb, whichever is greater
Detection Limit
Drift is negligible; detection limit is 1ppb
Response Time @ 90%
< 60 seconds
Calibration
Electrical zero, set automatically or chemical zero with hydrazine-free water. Calibration of the
slope compared with a laboratory instrument.
Analog Outputs
Two 4–20 mA/0–20 mA, linear or bilinear
Environmental
Enclosure
IP65/NEMA 4x
Storage Temperature Range
–20 to 60 °C (–4 to 140 °F)
Operating Temperature Range
5–45 °C (41–113 °F)
Relative Humidity
10 to 90% non-condensing
Compliance
UL listed (E226594) to UL3101-1& CSA C22.2 No. 1010.1