Modbus Communication
8.6 Network Connections
The 9186 uses RS485 digital communications. RS485 is multi-drop network,
meaning multiple devices can be “daisy chained” on a single pair of wires.
8.6.1 RS485 Connections
lists the connections to the RS485 network.
8.6.2 Terminator
To preventing “ringing” on the RS485 network, a terminator jumper is provided on
the RS485 interface board, next to connector K101. The terminator should be
enabled on the last device on the network. Set the jumper as follows:
Position 0
—Terminator disabled
Position 1
—Terminator enabled (last device on RS485 network)
8.6.3 Network Cable
The recommended wire is a two-wire shielded RS485 communication cable.
8.6.4 RS232 to RS485 Converter
An optional RS232 to RS485 converter may be required to convert the RS485
signal to RS232 on Programmable Logic Controllers (PLC) or Personal
Computers (PC). Only one converter is required for each RS485 network.
Table 23 RS485 Connections
Connector
Pin
Function
57012-00 Wire Color
K101
1
Data (+)
Blue
K101
2
Data (–)
White
K9
1 or 2
GND
Shield
Hach Catalog Number
Belden Part Number
Cable Type
57012-00
9841
2-wire Shielded
Hach Catalog Number
Description
57378-00
Isolated RS232 to RS485 converter, 115 V ac
57379-00
Isolated RS232 to RS485 converter, 230 V ac