Programming the Instrument
29
Figure 20
Linear and Bi-linear Output Loops
5.5 Setting Up the Display
Note:
The factory-defined
programming language is English.
The instrument will revert to English
after a software upgrade or when
the default values are restored.
1.
From the Main display, press the
RIGHT ARROW
key.
2.
Use the
LEFT ARROW
key to highlight SERVICE. Press
ENTER
.
3.
Use the
LEFT ARROW
key to highlight DISPLAY. Press
ENTER
.
4.
Program the parameters using
.
Table 7 mA Outputs Programming Parameters
Outputs 1/2
AFFECT
Conc. µA °C/°F
Used to determine whether the analog output is assigned to current, concentration, or
temperature measurement.
TYPE
0/20 or 4/20
Used to select the analog output range.
MODE
Lin or Bi-lin
Choose a linear or bi-linear scale (see
).
LOW
XXXX
Scale starting value
MIDDLE
XXXX
Scale middle value (bi-linear)
UP
XXXX
Scale ending value
Special Program
MODE
Preset/Last/Live
Action of the analog output during calibration, system alarm, maintenance, or active
timer cycles:
Return to a preset value
Return to the last value before the event
Live = live measurement
VALUE
XX
Used to define the return value (0–21 mA)
TEST
Used to test analog outputs in 1 mA steps (0–21 mA)
(%)
(mA)
Bi-linear
Linear
4
20
0
20
30
40
50
60
70
80
90 100
10