Oxygen analyser (ppb) - 9182 -
7-13
MALFUNCTION
POSSIBLE CAUSE
REMEDIES
Check to make sure that probe
is connected correctly
Initiate a software RESET
If steps 1-3 do not remedy the
problem, make a cold start
(RESET): interrupt the power
for 5-10 secs
Reprogram the instrument
again
Display reading static;
cannot be changed in
any way
Malfunctioning CPU board
and/or some other transmitter
hardware is defective
If problem persists, call the
Service Technician
Relays not energised
Instrument was programmed
incorrectly
Check whether the correct
relay parameters and setpoints
have been programmed
Hardware is defective
Check that the programmed
setpoints are compatible with
the programmed measuring
range
Check the relay characteristics
for proper functioning using an
ohmmeter
If problem persists, contact the
Service Technician
Instrument was incorrectly
programmed
Check the programmed output-
current parameters
Connection of the Monec with
peripherals (recorder, etc.)
are faulty, loose or defective
Check the cables
Compare the measured value
with the output current range
Wrong output current,
output current remains at
0 or 20 mA
Hardware is defective
If problem persists, call the
Service Technician
Wrong configuration
(3 electrode mode)
Check the swiches under the
amperometric board are
correctly positioned ("on")
Polarization voltage
incorrect
Wrong programming
Содержание POLYMETRON 9182
Страница 1: ...Operator Manual POLYMETRON 9182 Oxygen analyzer ppb 221 191 082 Revision G 20 11 2007 ...
Страница 2: ......
Страница 8: ......
Страница 39: ...Oxygen analyser ppb 9182 5 9 ...
Страница 48: ......
Страница 56: ......
Страница 70: ......
Страница 74: ......
Страница 76: ......
Страница 80: ......
Страница 86: ......
Страница 90: ...Oxygen analyser ppb 9182 A7 4 ...
Страница 91: ......
Страница 92: ......