Oxygen analyser (ppb) - 9182 -
5-5
5.1.3.2 ALARMS Menu
Relays S1...S4 may be allocated to the limit, alarm system or timer functions.
ò
LIMIT FUNCTION:
The alarm relays are activated if the comparison of the measured value with the limit programmed
meets the alarm function condition (up or down). The limits are programmed according to the
following programming variables:
ALARMS 1 … 4 (LIMIT)
AFFECT
- Conc.
- No
- °C/°F
Choice to use a limit on the measure, on
the temperature or not to use a limit.
LIM
xxxx
Enter a limit value
DIR.
- Up
- Down
Choice of the direction
DELAY
xxxs
Temporisation time before the relay is
commutated (in seconds)
HYST.
XX %
Definition of the hysteresis limit in %
(10 % max.)
The hysteresis operates only on one
side of the limit. The hysteresis is below
the limit for the high alarm (up) and
above the limit for the low alarm (down)
RELAY
- NO
- NC
Relay normally open or normally closed
Choix
F
ALARM 2
ALARM 3
ALARM 4
ALARMS
Select
ALARM 1
AFFECT. :
LIM. : 0.001 ppb
DIR : Down
DELAY : 000s
HYST. : 00 %
RELAY : NO
ALARMS
Select
Conc.
Содержание POLYMETRON 9182
Страница 1: ...Operator Manual POLYMETRON 9182 Oxygen analyzer ppb 221 191 082 Revision G 20 11 2007 ...
Страница 2: ......
Страница 8: ......
Страница 39: ...Oxygen analyser ppb 9182 5 9 ...
Страница 48: ......
Страница 56: ......
Страница 70: ......
Страница 74: ......
Страница 76: ......
Страница 80: ......
Страница 86: ......
Страница 90: ...Oxygen analyser ppb 9182 A7 4 ...
Страница 91: ......
Страница 92: ......