Oxygen analyser (ppb) - 9182 -
5-10
OUTPUTS 1/2
AFFECT
- Conc
- µA
- °C/°F
Choice of the analogue output allocation to the cell current, to
the concentration or to the temperature measurement
TYPE
0/20
4/20
Choice of the analogue output type
MODE
- Lin
- Dual
Choice between a linear or dual range (see drawing below)
LOWER
XXXX
Bottom of the scale value
MIDD.
XXXX
Mid-scale value (only in
dual mode
)
UPPER
XXXX
Top of the scale value
SPECIAL PROG.
MODE
- last
- preset
- live
Characteristics of the analogue output during calibration, alarm
system, maintenance or timer active cycles: frozen to the latest
stored before any operation listed above, forced to a preset
value, live measurement
VALUE
XX
Preset value (0 to 21 mA)
TEST
Test the analogue outputs by steps of 1 mA (0...21 mA)
Illustration of the linear/dual outputs scales:
(%)
Upper
20
20
16
15
Middle
12
10
8
5
Lower
4
0
0
10
20
30
40
50
60
70
80
90 100
(mA)
Dual
Linear
Содержание POLYMETRON 9182
Страница 1: ...Operator Manual POLYMETRON 9182 Oxygen analyzer ppb 221 191 082 Revision G 20 11 2007 ...
Страница 2: ......
Страница 8: ......
Страница 39: ...Oxygen analyser ppb 9182 5 9 ...
Страница 48: ......
Страница 56: ......
Страница 70: ......
Страница 74: ......
Страница 76: ......
Страница 80: ......
Страница 86: ......
Страница 90: ...Oxygen analyser ppb 9182 A7 4 ...
Страница 91: ......
Страница 92: ......