Calibration Results

A calibration screen will be displayed showing
current measurement data which is continually
refreshed. 
The value “% ideal current” is a percentage of the
cu rren t ag ainst  t he  id eal curren t fo r th e
membrane type selected. If this percentage is not
within the accepted range, an error message is
displayed and the calibration process fails. A
warning message can be displayed when this
value is close to the boundaries, but when
calibration can be accepted.
The message is first displayed in the result box.
The dialog box with the error message or the
warning is displayed when the finish button is
pressed.
The value “% last calibration” shows the ratio
between current measurement and the previous
sensor calibration. 
The value “% variation” indicates the variation
during the last 3 measurements, which is the
stability of the measurements. A variation as low
as possible is needed for a precise calibration 
The display shows the actual calibration
p a r a m e t e r s ,   a n d   t h e   a c t u a l   r e a d i n g s
(temperature, barometer, current).

In case of a calibration failure, consider replacing
the membrane. See the EC Sensor Maintenance
Manual for details.

Verification

Similar to the calibration procedure, but for
verification of the actual calibration values. The
result of the measurements made during the
verification is not stored and the actual calibration
data is not modified.

O

2

 Sensor Calibration

The sensor is in contact with either:

Air at atmospheric pressure (In Air)

O

2

 at known concentration (Direct value). 

O

2

 In Air Calibration

This calibration procedure places the O

2

 sensor in

water-saturated air, to provide a known oxygen
reference against which to calibrate.
Dry the sensor thoroughly, before placing the
sensor storage cap under tap water. Shake off
any excess water, but leave a few drops inside
the cap. Verify that the screw-on protection cap is
in place on the sensor head. 
Then, loosely place the storage cap back on the
sensor, holding it in place with a few turns of its
collar.
Set the calibration parameters accordingly and
press calibrate.

O

2

 Direct Calibration

This procedure calibrates the oxygen sensor
against a liquid sample containing a known level
of dissolved O

2

 flowing through the sample line.

The instrument displays the sensitivity of the
sensor as a percentage of the sensitivity
determined when calibration was last performed. 

Set the calibration parameters accordingly and
press calibrate.

Calibration (cont)

Содержание ORBISPHERE 511

Страница 1: ...Operator Quick Guide ORBISPHERE 510 511 512 Revision G 05 11 2008...

Страница 2: ...d service routines can be called by pressing buttons and menus bar on screen Measurement display shows one measurement slope per sensor in function up to 3 lines for a 3 channel Orbisphere 51x Special...

Страница 3: ...s measured and the channel configuration The statistical view offers statistical data that matches with Total Quality management tools Statistics is a tool to better analyze how a process behaves The...

Страница 4: ...concentration is smaller than that entered in Variation Time The stop criteria is met when the elapsed time reaches the Max time parameter The parameter Max time is the maximum time allowed to reach...

Страница 5: ...arms Set this to a minimum but enough to avoid alarms for non representative peaks beyond the set level Measurement Filter Configuration The filters are aimed at flattening the measurement curve in si...

Страница 6: ...actor Enable correction as appropriate If checked the percentage correction factor must be entered This value cannot be negative Note To manually set the TC sensor into a continuous purge mode press t...

Страница 7: ...from the non volatile memory Adjust the parameters for recording and storing measured data Two storage mode selections are available depending on the measurement mode selected Continuous measurement m...

Страница 8: ...ation of the Measured Gas Before initiating a calibration process the calibration parameters must be set by pressing on the Modify button Last calibration parameters are memorized so this step can be...

Страница 9: ...intenance Manual for details Verification Similar to the calibration procedure but for verification of the actual calibration values The result of the measurements made during the verification is not...

Страница 10: ...the Measured Gas Before initiating a calibration process the calibration parameters must be set by pressing on the Modify button The last calibration parameters are memorized so this step can be igno...

Страница 11: ...e variation indicates the variation during the last 3 measurements which is the stability of the measurements A variation as low as possible is needed for a precise calibration The display shows the a...

Страница 12: ...iate the H2 interference calibration Interference Calibration TC Sensor This calibration is available for TC sensors and is required when one of the interferences has been enabled The lower part of th...

Страница 13: ...point calibration recommended One point calibration By default the two point calibration is selected Two Point Calibration recommended Connect a certified absolute pressure gauge to the sample line a...

Страница 14: ...e information on this please refer to the manual s delivered with your sensor s Language Check the language as required and restart the instrument to apply the change The instrument will restart in th...

Страница 15: ...tration is below the Alarm LowLow limit Alarm low The gas concentration is below the Alarm Low limit Alarm high The gas concentration is above the Alarm High limit Alarm high high The gas concentratio...

Страница 16: ......

Отзывы: