Hach Ultra ORBISPHERE 511 Скачать руководство пользователя страница 10

Page 10

O

3

 Sensor Calibration

The sensor is either in contact with:

Air at atmospheric pressure (In Air)

O

3

 at known concentration (Direct Value). 

The procedure is the same as for the O

2

 sensor.

In the case of the "In air" calibration, the sensor
me asu re s  O

2

 d uring  calibrat ion.  Th e  O

3

coefficient is deduced taking into account how the
sensor behaves in O

2

As a different voltage is used at the anode to
measure O

2

 and O

3

, the O

3

 measurement takes a

long time to stabilize. To facilitate the follow up
after an "O

3

 in air" calibration, negative values

can be displayed.

H

2

 Sensor Calibration

T h e   r e c o m m e n d e d   m e t h o d   i s   a t   k n o w n
concentration (Direct value). The gas can be pure
H

2

 or a mixture of H

2

 with an inert gas (e.g. a

mixture of H

2

/N

2

). The known concentration is

entered by the user in the calibration parameter
window. The sensor is in contact with calibration
gas (Direct Value) in gas phase at atmospheric
pressure. 
Make sure the H

2

 concentration used for

calibration is within the acceptable range for the
membrane.

TC Gas Sensor Calibration

Calibration of the Measured Gas

Before initiating a calibration process, the
calibration parameters must be set by pressing on

t h e  

M o d i f y

  b u t t o n .   T h e   l a s t   c a l i b r a t i o n

parameters are memorized, so this step can be
ignored if the correct parameters are already set. 
Similarly, if only the calibration value has
changed, then this can be updated directly
instead of pressing the 

Modify

 button.

Modify Calibration Parameters

Gas Phase: Select 

liquid

 or 

gas 

(direct 

calibration only)

Gas unit type: 

Partial

fraction

 or 

dissolved

(dissolved is for calibration in a liquid 
only)

Gas unit: The list of available units 
depends on unit type selected above.

Liquid: Select as appropriate.

Enter the gas concentration according to 
the value in the calibration media.

Hold during calibration: On by default, 
this stops any output from the instrument 
during the calibration process to avoid 
sending invalid information to any 
connected device. 

Automatic calibration stop: If selected, 
when the stability criteria is reached, the 
calibration process stops automatically.

Interferences enabled: If selected, this 
takes into account the influence of 
interferences during calibration. By 
default the same interferences as during 
measurement are selected.

Press OK to start calibration

Calibration (cont)

Содержание ORBISPHERE 511

Страница 1: ...Operator Quick Guide ORBISPHERE 510 511 512 Revision G 05 11 2008...

Страница 2: ...d service routines can be called by pressing buttons and menus bar on screen Measurement display shows one measurement slope per sensor in function up to 3 lines for a 3 channel Orbisphere 51x Special...

Страница 3: ...s measured and the channel configuration The statistical view offers statistical data that matches with Total Quality management tools Statistics is a tool to better analyze how a process behaves The...

Страница 4: ...concentration is smaller than that entered in Variation Time The stop criteria is met when the elapsed time reaches the Max time parameter The parameter Max time is the maximum time allowed to reach...

Страница 5: ...arms Set this to a minimum but enough to avoid alarms for non representative peaks beyond the set level Measurement Filter Configuration The filters are aimed at flattening the measurement curve in si...

Страница 6: ...actor Enable correction as appropriate If checked the percentage correction factor must be entered This value cannot be negative Note To manually set the TC sensor into a continuous purge mode press t...

Страница 7: ...from the non volatile memory Adjust the parameters for recording and storing measured data Two storage mode selections are available depending on the measurement mode selected Continuous measurement m...

Страница 8: ...ation of the Measured Gas Before initiating a calibration process the calibration parameters must be set by pressing on the Modify button Last calibration parameters are memorized so this step can be...

Страница 9: ...intenance Manual for details Verification Similar to the calibration procedure but for verification of the actual calibration values The result of the measurements made during the verification is not...

Страница 10: ...the Measured Gas Before initiating a calibration process the calibration parameters must be set by pressing on the Modify button The last calibration parameters are memorized so this step can be igno...

Страница 11: ...e variation indicates the variation during the last 3 measurements which is the stability of the measurements A variation as low as possible is needed for a precise calibration The display shows the a...

Страница 12: ...iate the H2 interference calibration Interference Calibration TC Sensor This calibration is available for TC sensors and is required when one of the interferences has been enabled The lower part of th...

Страница 13: ...point calibration recommended One point calibration By default the two point calibration is selected Two Point Calibration recommended Connect a certified absolute pressure gauge to the sample line a...

Страница 14: ...e information on this please refer to the manual s delivered with your sensor s Language Check the language as required and restart the instrument to apply the change The instrument will restart in th...

Страница 15: ...tration is below the Alarm LowLow limit Alarm low The gas concentration is below the Alarm Low limit Alarm high The gas concentration is above the Alarm High limit Alarm high high The gas concentratio...

Страница 16: ......