Interferences Configuration (TC Sensor)

A number of different interfering gas options are
available (temperature, H

2

O, He, O

2

 and H

2

),

depending on the instrument configuration and
the gas being measured. 

Temperature compensation: If there is a 
significant fluctuation in the sample 
temperature enable temperature 
compensation to reduce the influence of 
these temperature fluctuations.

Interfering H2O: This only applies to 
measurements taken in 100% humid 
gases or in traces of dissolved gas in 
water. It allows a correction to the slope 
and temperature coefficients calculated 
during calibration.

Interfering gas: Enable this option to 
allow the instrument to correct for the 
presence of specific gases in the sample. 
If the interfering gas is helium, you can 
enter the concentration percentage if 
known.

Measured Data Storage 

There is one measurement file per channel which
contains the data generated by the measurement

cycle. The current measurement files are updated
in volatile memory, and are regularly copied in
non-volatile memory (file back-up). At start up, the
measurement files in volatile memory are updated
with the files from the non volatile memory.

Adjust the parameters for recording and storing
measured data. Two storage mode selections are
available depending on the measurement mode
selected:
Continuous measurement mode:

No storage

Store once: When volatile memory is full 
(10,000 positions), measurement 
recording stops

Rolling buffer: When volatile memory is 
full, the latest measurement set replaces 
the oldest (first-in, first-out basis)

Sample measurement mode:

Only final measurement: The 
measurement when the stop criteria are 
met is saved (one per sample)

Only when sample mode started: The 
measurements when the sample mode is 
started are saved

All measurements: All measurements 
stored continuously

Additional parameters:

RAM time (volatile memory): Delay in 
seconds between two recordings of 
measured data.

FLASH time (non-volatile memory): 
Delay in seconds between two data file 
transfers from volatile memory into non-
volatile memory. Last data file erases 
previous one.

Buttons:

Purge data: Clear all data in the volatile 
and non-volatile memories

Start logging measurement: In store once 
mode, starts and stops the measurement 
recording session. Measurement 
recording is stopped when buffer is full

Open data 1,2, or 3 (and TPO or TPA 
data if enabled): Opens the table 
showing the measured value which are 
stored in the volatile memory (RAM)

Measurement (cont)

Содержание ORBISPHERE 511

Страница 1: ...Operator Quick Guide ORBISPHERE 510 511 512 Revision G 05 11 2008...

Страница 2: ...d service routines can be called by pressing buttons and menus bar on screen Measurement display shows one measurement slope per sensor in function up to 3 lines for a 3 channel Orbisphere 51x Special...

Страница 3: ...s measured and the channel configuration The statistical view offers statistical data that matches with Total Quality management tools Statistics is a tool to better analyze how a process behaves The...

Страница 4: ...concentration is smaller than that entered in Variation Time The stop criteria is met when the elapsed time reaches the Max time parameter The parameter Max time is the maximum time allowed to reach...

Страница 5: ...arms Set this to a minimum but enough to avoid alarms for non representative peaks beyond the set level Measurement Filter Configuration The filters are aimed at flattening the measurement curve in si...

Страница 6: ...actor Enable correction as appropriate If checked the percentage correction factor must be entered This value cannot be negative Note To manually set the TC sensor into a continuous purge mode press t...

Страница 7: ...from the non volatile memory Adjust the parameters for recording and storing measured data Two storage mode selections are available depending on the measurement mode selected Continuous measurement m...

Страница 8: ...ation of the Measured Gas Before initiating a calibration process the calibration parameters must be set by pressing on the Modify button Last calibration parameters are memorized so this step can be...

Страница 9: ...intenance Manual for details Verification Similar to the calibration procedure but for verification of the actual calibration values The result of the measurements made during the verification is not...

Страница 10: ...the Measured Gas Before initiating a calibration process the calibration parameters must be set by pressing on the Modify button The last calibration parameters are memorized so this step can be igno...

Страница 11: ...e variation indicates the variation during the last 3 measurements which is the stability of the measurements A variation as low as possible is needed for a precise calibration The display shows the a...

Страница 12: ...iate the H2 interference calibration Interference Calibration TC Sensor This calibration is available for TC sensors and is required when one of the interferences has been enabled The lower part of th...

Страница 13: ...point calibration recommended One point calibration By default the two point calibration is selected Two Point Calibration recommended Connect a certified absolute pressure gauge to the sample line a...

Страница 14: ...e information on this please refer to the manual s delivered with your sensor s Language Check the language as required and restart the instrument to apply the change The instrument will restart in th...

Страница 15: ...tration is below the Alarm LowLow limit Alarm low The gas concentration is below the Alarm Low limit Alarm high The gas concentration is above the Alarm High limit Alarm high high The gas concentratio...

Страница 16: ......

Отзывы: