Operator Quick Guide
ORBISPHERE 510/511/512
Revision G - 05/11/2008
Страница 1: ...Operator Quick Guide ORBISPHERE 510 511 512 Revision G 05 11 2008...
Страница 2: ...d service routines can be called by pressing buttons and menus bar on screen Measurement display shows one measurement slope per sensor in function up to 3 lines for a 3 channel Orbisphere 51x Special...
Страница 3: ...s measured and the channel configuration The statistical view offers statistical data that matches with Total Quality management tools Statistics is a tool to better analyze how a process behaves The...
Страница 4: ...concentration is smaller than that entered in Variation Time The stop criteria is met when the elapsed time reaches the Max time parameter The parameter Max time is the maximum time allowed to reach...
Страница 5: ...arms Set this to a minimum but enough to avoid alarms for non representative peaks beyond the set level Measurement Filter Configuration The filters are aimed at flattening the measurement curve in si...
Страница 6: ...actor Enable correction as appropriate If checked the percentage correction factor must be entered This value cannot be negative Note To manually set the TC sensor into a continuous purge mode press t...
Страница 7: ...from the non volatile memory Adjust the parameters for recording and storing measured data Two storage mode selections are available depending on the measurement mode selected Continuous measurement m...
Страница 8: ...ation of the Measured Gas Before initiating a calibration process the calibration parameters must be set by pressing on the Modify button Last calibration parameters are memorized so this step can be...
Страница 9: ...intenance Manual for details Verification Similar to the calibration procedure but for verification of the actual calibration values The result of the measurements made during the verification is not...
Страница 10: ...the Measured Gas Before initiating a calibration process the calibration parameters must be set by pressing on the Modify button The last calibration parameters are memorized so this step can be igno...
Страница 11: ...e variation indicates the variation during the last 3 measurements which is the stability of the measurements A variation as low as possible is needed for a precise calibration The display shows the a...
Страница 12: ...iate the H2 interference calibration Interference Calibration TC Sensor This calibration is available for TC sensors and is required when one of the interferences has been enabled The lower part of th...
Страница 13: ...point calibration recommended One point calibration By default the two point calibration is selected Two Point Calibration recommended Connect a certified absolute pressure gauge to the sample line a...
Страница 14: ...e information on this please refer to the manual s delivered with your sensor s Language Check the language as required and restart the instrument to apply the change The instrument will restart in th...
Страница 15: ...tration is below the Alarm LowLow limit Alarm low The gas concentration is below the Alarm Low limit Alarm high The gas concentration is above the Alarm High limit Alarm high high The gas concentratio...
Страница 16: ......