4.
Select 1 point cal.
5.
Select the output mode.
Option
Description
Active
Use the current measured value during a calibration
Hold
Use the last measured value during a calibration
Transfer
Transfer a specified value to the 4–20 mA input for use during a
calibration
6.
Use the arrow keys to edit the displayed value so it matches the
input value from the outside device as measured by an accurate
current meter or as otherwise determined. Push the
ENTER
key.
The display shows the Current, Input and Cal values.
7.
Push the
ENTER
key to continue.
If the calibration passes, the display shows a Calibration passed
message, along with the Offset and Slope values. Push the
ENTER
key to continue and return to the Calibrate menu screen.
If the calibration fails, the display shows a Calibration failed
message. Refer to
Troubleshooting calibration failure
on page 9
for troubleshooting suggestions.
To repeat the calibration, push the
ENTER
key.
To restore the previous calibration, push the
BACK
key. Then push
the
ENTER
key or wait for the controller to return to the Calibrate
menu screen.
Perform a 2 point calibration
1.
From the Settings Menu, select Sensor Setup.
2.
Select Calibrate.
3.
Select Cal Options.
a.
Select the Current option to perform a calibration using current
(mA) values, or
b.
Select the Input option to perform a calibration using
measurement values, such as psi.
c.
Push the
ENTER
key.
4.
Select 2 point cal.
5.
Select the output mode.
Option
Description
Active
Use the current measured value during a calibration
Hold
Use the last measured value during a calibration
Transfer
Transfer a specified value to the 4–20 mA input for use during a
calibration
6.
Use the arrow keys to edit the point 1 value.
Note: The displayed point 1 value must match the point 1 value from the
outside device as measured by an accurate current meter or as otherwise
determined.
7.
Push the
ENTER
key to continue.
The display shows the Current, Input and Cal values for cal point 1.
8.
Push the
ENTER
key to continue.
9.
Use the arrow keys to edit the point 2 value. The point 2 value should
be as far as possible from the point 1 value for better 2 point
calibration.
Note: The displayed point 2 value must match the point 2 value from the
outside device as measured by an accurate current meter or as otherwise
determined.
10.
Push the
ENTER
key to continue. The display shows the Current,
Input and Cal values for cal point 2.
11.
Push the
ENTER
key to continue.
If the calibration passes, the display shows a Calibration passed
message, along with the Offset and Slope values. Push the
ENTER
key to continue and return to the Calibrate menu screen.
If the calibration fails, the display shows a Calibration failed
message. Refer to
Troubleshooting calibration failure
on page 9
for troubleshooting suggestions.
To repeat the calibration, push the
ENTER
key.
To restore the previous calibration, push the
BACK
key. Then push
the
ENTER
key or wait for the controller to return to the Calibrate
menu screen.
8
English
Содержание sc200 4-20
Страница 5: ...English 5 ...
Страница 6: ...6 English ...
Страница 12: ...12 Deutsch ...
Страница 13: ...Deutsch 13 ...
Страница 20: ...20 Italiano ...
Страница 21: ...Italiano 21 ...
Страница 28: ...28 Français ...
Страница 29: ...Français 29 ...
Страница 36: ...36 Español ...
Страница 37: ...Español 37 ...
Страница 44: ...44 Português ...
Страница 45: ...Português 45 ...
Страница 51: ...Čeština 51 ...
Страница 52: ...52 Čeština ...
Страница 58: ...58 Dansk ...
Страница 59: ...Dansk 59 ...
Страница 65: ...Nederlands 65 ...
Страница 66: ...66 Nederlands ...
Страница 72: ...72 Polski ...
Страница 73: ...Polski 73 ...
Страница 79: ...Svenska 79 ...
Страница 80: ...80 Svenska ...
Страница 86: ...86 Suomi ...
Страница 87: ...Suomi 87 ...
Страница 94: ...94 български ...
Страница 95: ...български 95 ...
Страница 102: ...102 Magyar ...
Страница 103: ...Magyar 103 ...
Страница 109: ...Română 109 ...
Страница 110: ...110 Română ...
Страница 117: ...lietuvių kalba 117 ...
Страница 118: ...118 lietuvių kalba ...
Страница 124: ...124 Русский ...
Страница 125: ...Русский 125 ...
Страница 132: ...132 Türkçe ...
Страница 133: ...Türkçe 133 ...
Страница 139: ...Slovenský jazyk 139 ...
Страница 140: ...140 Slovenský jazyk ...
Страница 146: ...146 Slovenski ...
Страница 147: ...Slovenski 147 ...
Страница 153: ...Hrvatski 153 ...
Страница 154: ...154 Hrvatski ...
Страница 160: ...160 Ελληνικά ...
Страница 161: ...Ελληνικά 161 ...
Страница 168: ...168 eesti keel ...
Страница 169: ...eesti keel 169 ...
Страница 174: ......
Страница 175: ......