10.4 Disposal
Electrical and electronic devices must be disposed of separately from
household waste! This device was made available for sale after the
13th August 2005. For correct disposal, please contact your Haag-
Streit representative. This will guarantee that no hazardous sub-
stances enter the environment and that valuable raw materials are re-
cycled.
10.5 Observed standards
EN 60601-1
EN ISO 15004-1
EN 60601-1-2
EN ISO 15004-2
EN 62471
EN ISO 10343
EN 60825-1
EN ISO 22665
EN 60529
EN ISO 19980
ANSI Z80.23
ANSI Z80.36
10.6 Information and manufacturerÕs declaration concerning
electromagnetic compatibility (EMC)
10.6.1 General
This device fulfills the requirements on electromagnetic compatibility according to
IEC 60601-1-2:2014 (4th Edition). The device is built so that the generation and
emission of electromagnetic interference is limited to the extent that other devices
are not disturbed in their use in accordance with the regulations and so that the
device itself is suitably immune to electromagnetic interference.
WARNING!
• Electrical medical devices and systems are subject to special EMC
measures and must be installed in accordance with the EMC in-
structions contained in this accompanying document.
• Use of accessories, transducers and cables other than those spe-
cified or provided by Haag-Streit could result in increased electro-
magnetic emissions or decreased electromagnetic immunity of this
device and result in improper operation.
• Third-party devices may only be connected in compliance with the
IEC 60601-1 standard.
WARNING!
• Avoid damages due to high electrostatic discharges (ESD). Electro-
static discharges with voltages exceeding 6 kV to some parts of the
system like joystick or metallic parts on the instrument base may in-
fluence the device.
• The communication between the Lenstar 900 and the PC could be
interrupted; this requires a restart of the EyeSuite software and a re-
petition of the investigation.