7.4.7 Pupillometry

Measurement range:

2 – 13 mm

Display resolution:

0.01 mm

The above mentioned measurement ranges correspond to the default setting for the automatic ana-
lysis.
The in vivo reproducibility was evaluated in a clinical study of cataract patients (see table 'All eyes'
and table 'Special eyes')
1 σ = One standard deviation

7.4.8 Study design

• The clinical trial for deriving in vivo reproducibility was approved by the local eth-

ics committee. The study was planned and conducted as a prospective, non-

randomised comparative study.

• Two phases were scheduled in the approved study protocol. In the first phase,

measurements of axial length (AL), central corneal thickness (CCT), anterior

chamber depth (ACD), central lens thickness (LT), median corneal radius (R)

and the axis position of the flat meridian (axis) were taken.

• The white-to-white distance (WTW) was measured in the second phase.

• A total of 144 eyes in 80 subjects were included in study phase 1 and a total of

40 eyes in 20 subjects in study phase 2.

• Subjects with different conditions of the anterior and posterior segment of the

eye (cataract in different stages, pseudophakia with different IOLs, aphakia, sil-

icone-oil fill) and subjects with healthy eyes were included in this clinical study.

Data were analyzed for all eyes (see table 'All eyes') and for a subgroup of eyes

with special health conditions (see table 'Special eyes'), with a complete set of 5

repeat measurements on both eyes of each subject.

• The special eyes group included eyes with one or more of the following condi-

tions: pseudophakia, aphakia and silicone-oil fill.

Tab. 1: All eyes

[unit]

n

Mean

grand

SD

repeat

CV

AL [mm]

45 / 90

23.973

0.035

0.00145

CCT [μm]

53 / 106

557.1

2.3

0.00407

ACD [mm]

34 / 68

3.19

0.04

0.01220

LT [mm]

27 / 54

4.56

0.08

0.01784

R [mm]

34 / 68

7.67

0.03

0.00396

Axis [°]

27 / 54

72

11

0.14191

WTW [mm]

9 / 18

12.27

0.04

0.00337

Tab. 2: Special eyes

[unit]

n

Mean

grand

SD

repeat

CV

AL [mm]

10 / 20

24.087

0.056

0.00234

CCT [μm]

11 / 22

564.4

2.8

0.00496

ACD [mm]

5 / 10

7.75

0.03

0.00333

Axis [°]

3 / 6

80

13

0.16092

Abbreviations

n

Number of subjects / number of eyes

Mean

grand

Overall mean of results on all eyes

SD

repeat

Repeatability standard deviation

CV

Coefficient of variation

SD

Standard deviation

LT

Lens thickness

AL

Axial length

R

Corneal radius of curvature

CCT

Central corneal thickness

Axis

Axis of flat meridian

ACD

Anterior chamber depth

WTW

White-to-white distance

Содержание LENSTAR LS 900

Страница 1: ...SCH ENGLISH FRAN AIS ITALIANO ESPA OL NEDERLANDS PORTUGU S SVENSKA HAAG STREIT AG 3098 Koeniz Switzerland HS Doc no 1500 7220055 04150 2022 09 INSTRUCTIONS FOR USE Biometer Lenstar LS 900 15 Edition 2...

Страница 2: ...for use 4 1 2 Ambient conditions 4 1 3 Shipment and unpacking 4 1 4 Installation warnings 4 1 5 Operation environment 5 1 6 Plausibility of the measurements 6 1 7 IOL calculation 7 1 7 1 References 7...

Страница 3: ...4 Measured variables Lenstar LS 900 17 7 4 1 Central corneal thickness CCT 17 7 4 2 Anterior chamber depth ACD 17 7 4 3 Lens thickness LT 17 7 4 4 Axial length AL 17 7 4 5 Keratometry R 17 7 4 6 Whit...

Страница 4: ...ts This report must be signed by you and by the representative of the transport company Leave the device in the packaging for a few hours before unpacking it condens ation Check the device for damage...

Страница 5: ...ipment including peri pherals such as antenna cables and external antennas should be used no closer than 30 cm 12 inches to any part of the device in cluding cables specified by Haag Streit Otherwise...

Страница 6: ...ake into account the type e g posterior subcapsular cataract and density of the cataract when evaluating plausibility Prior to the measurement the user must verify that the patient is not wearing cont...

Страница 7: ...constant When using the Lenstar LS 900 only IOL constants optimised for op tical biometers should be used Please contact your IOL manufacturer for informa tion on optimised IOL constants for optical b...

Страница 8: ...ividual meas urements are performed per day and per patient s eye NOTE In accordance with EN 60825 1 the limiting values for class 1 lasers are respected when the device is used in the defined manner...

Страница 9: ...ns thickness Corneal curvature Radii for flat and steep meridian Axis of the flat meridian White to white distance Pupil diameter 2 1 Device description The Lenstar LS 900 is a non contact device for...

Страница 10: ...s must be at least 6 years old 2 2 4 Contraindications There are no known contraindications 2 3 Principles of operation The device uses contact less measurement technologies OLCR interferometry to mea...

Страница 11: ...5 Control lever 6 Type plate 7 Side identification sticker Headrest option 8 Headrest 9 Head band 10 Mark for optimum eye height 11 Chin rest 12 Chin rest height adjustment 13 Hand grips for patient T...

Страница 12: ...supplied USB cable 2 m Connect the electric power supply cable Integral mains components work with the voltages specified in section Technical data under chapter Primary side It is not necessary to se...

Страница 13: ...suring with Lenstar APS optional WARNING The Lenstar APS Automated Positioning System device moves autonomously during the automated measurement process To avoid trapping fingers do not touch any movi...

Страница 14: ...Cleaning and disinfection Make sure the T Cone is dry before measuring Retract the Lenstar T Cone completely before positioning the patient in the headrest Always start the measuring procedure with th...

Страница 15: ...nged in two rings with the following diameters standard eye R 7 8 mm Outer measuring points Inner measuring points 2 3 mm 1 65 mm For each measuring point the equivalent of an ideal sphere is calculat...

Страница 16: ...are on the PC Switch off the PC Disconnect the power supply plug from the mains if you do not intend to use the device for an extended period of time 7 Technical data Type designation LS 900 Dimension...

Страница 17: ...sured variables Lenstar LS 900 Measurement ranges are based on the Phakic measurement mode 7 4 1 Central corneal thickness CCT Measurement range 300 800 m Display resolution 1 m In vivo repeatability...

Страница 18: ...with different conditions of the anterior and posterior segment of the eye cataract in different stages pseudophakia with different IOLs aphakia sil icone oil fill and subjects with healthy eyes were...

Страница 19: ...R1 7 987 R2 7 584 0 006 0 065 0 66 0 004 0 044 0 33 Ellipsoid mm R 7 79 k 0 255 0 001 0 083 0 78 0 020 0 038 1 51 Sphere 1 mm R 6 448 0 013 0 037 0 61 0 012 0 025 1 12 Sphere 2 mm R 7 804 0 008 0 034...

Страница 20: ...user s disinfection process Comply with the stipulated exposure time Observe the manufacturer s safety instructions Too strong or aggressive disinfectants or cleaning liquids e g hydro gen peroxide wi...

Страница 21: ...ins cable USA 7220761 Mains cable UK 7220762 Upgrade Kit LS 900 7225044 Headrest Please refer to the instructions for use of headrest 10 2 Legal regulations Haag Streit maintains a quality management...

Страница 22: ...on and emission of electromagnetic interference is limited to the extent that other devices are not disturbed in their use in accordance with the regulations and so that the device itself is suitably...

Страница 23: ...F emissions CISPR 11 Group 1 This product uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipm...

Страница 24: ...applicable for DC and I O if cable 3 m Surge IEC 61000 4 5 0 5 1 kV line s to line s 0 5 1 2 kV line s to earth 1 kV line s to line s 0 5 1 2 kV line s to earth Mains power quality should be that of a...

Страница 25: ...radio amateur band 3 Vrms 150 kHz 80 MHz outside ISM bands and radio amateur band 6 Vrms 150 kHz 80 MHz in ISM bands and radio amateur band If the measured field strength in the location in which thi...

Страница 26: ...vicinity of equipment marked with the following symbol The ISM industrial scientific and medical bands between 150 kHz and 80 MHz are 6 765 6 795 MHz 13 553 13 567 MHz 26 957 27 283 MHz 40 66 40 7 MH...

Страница 27: ...kHz 80 MHz outside ISM and ra dio amateur bands d 1 2 P 150 kHz 80 MHz in ISM and radio am ateur bands d 2 0 P 80 MHz 6 GHz for define RF Wire less transmitters see table before d 1 2 P 0 01 0 12 0 2...

Страница 28: ...ould you have any further questions please contact your Haag Streit representative at www haag streit com haag streit group contact haag streit distributors distributors HAAG STREIT AG Gartenstadtstra...

Отзывы: