4 Device assembly / installation
WARNING!
• Do not modify this device without authorization of the manufacturer.
Installation and repairs may only be performed by trained
specialists.
• Contact your Haag-Streit representative for installation, repairs and
modification work on the system. The contact details are available at
www.haag-streit.com.
• Only original Haag-Streit spare parts may be used.
4.1 Microscope and illumination
• The slit lamp is packaged and delivered fully assembled. The transport safety
devices must be removed prior to commissioning.
• Fix the breath shield (35) in place by fastening the knurled screw (34) on the
inside of the bearer arm.
4.2 Power supply
NOTE!
• Observe the respective Haag-Streit instructions for use. For further
information, please contact your Haag-Streit representative.
• This device must only be operated with PS-LED and PS-LED HSM
901 Haag-Streit power supplies and the RM02 release module.
4.3 Instrument base with weight compensation facility
The weight of additional accessories mounted on the microscope can be
compensated using counterbalance springs. This keeps the height adjustment of the
slit lamp easy.
4.4 Setting the weight compensation facility
Turn the control lever (60) to its lowest position and then loosen it slightly a quarter
turn. Turn the microscope and illumination to the side. Apply 1−3 springs depending
on the accessory.
4.5 Switching on the compensation facility
Turn anticlockwise until the screws (59) are completely released.
4.6 Switching off the compensation facility
Turn clockwise until you meet resistance. Verify whether the microscope arm
springs back downwards if you push it upwards with your hand. This will only
happen if the load is already at maximum. Generally, as many counterbalance
springs should be deactivated as necessary until this spring action occurs. The
weight compensation facility is set correctly once the illumination and microscope
with the mounted accessories weigh slightly more than the counterbalance springs.
4.7 Regulating the clearance of the slit width controls
The small screw in the centre of the right control knob
(83) allows you to regulate the friction of the turning
movement of these adjusting knobs. Turning it slightly
to the right (in) makes it harder, turning it left (out)
makes it easier. It should at least be set so hard that
the slit cannot close on its own.
5 Commissioning
The device can be switched on and off using the mains power switch on the power
supply. The green lamp in the rocker switch lights up when the device is switched
on.
5.1 Switching on the device
• Connect the power supply to the mains and press the rocker switch. The green
lamp in the rocker switch lights up when the device is switched on.
• Turn the rotating knob on the illumination control to a position between '1' and
'10'.