• Insert the test rod into the radial movement bearing, whilst at the same time
aligning the surface to the microscope at a right angle.
• Set the slit length to 8 or 14 mm.
• Set the illumination intensity to 50%.
• Set the magnification in the microscope to max.
• Set the eyepieces in such a way that the test rod is in sharp focus. In doing so,
turn the ocular from the (+) to the (-) side.
• For all magnifications, the structure of the test rod must be shown in sharp
focus.
• Close the slit edges to approx. 0.5 mm. The edges must be shown in sharp
focus.
• Open the slit edges completely and turn the test rod by 45°, the sharp area must
be in the centre of the test rod.
9.2 Servicing
To guarantee a long service life, the device must be cleaned weekly as described
and protected with the dust cover when not in use. We recommend having the
device inspected once a year by an authorized service technician.
9.3 Cleaning and disinfection
The Haag-Streit slit lamps and their accessories can, if required, be carefully wiped
down with ready-for-use disposable 70% ethanol disinfectant wipes. Surface-friendly
disinfectants (containing aldehyde or aldehyde-free) are also permitted, such as
Kohrsolin FF.
WARNING!
• The preparation instructions provided do not apply to tonometer
measuring prisms.
• Tonometer measuring prisms must be prepared in accordance with
a different manual.
• Too strong or aggressive disinfectants or cleaning liquids e.g.
hydrogen peroxide will damage the finish and coating of the device.
• Do not use sprays.
• Observe the manufacturer's safety instructions.
• Do not use any cloths that drip.
• Wring out any soaked cloths before use when necessary.
• Ensure that no liquid penetrates the device.
• Comply with the stipulated exposure time.
• Clean optical surfaces after disinfection with a very soft cloth.
NOTE!
IP code: IPX0 (device is not protected against liquids)
9.4 Replacing the illumination mirror
The mirror can be most easily accessed if the microscope is turned away from the
illumination and the illumination inclined two points.
WARNING!
Only use mirrors with a LOT number.
9.5 Dust cover
We recommend protecting the slit lamp with a dust cover when
not in use.
10 Appendix
WARNING!
• Do not modify this device without authorization of the manufacturer.
Installation and repairs may only be performed by trained
specialists.
• Contact your Haag-Streit representative for installation, repairs and
modification work on the system. The contact details are available at
www.haag-streit.com.
• Only original Haag-Streit spare parts may be used.