Diffuse illumination:
• Connecting the diffusor (14) upstream creates diffused illumination. This allows
monitoring of the overview and can be used to capture the overview with an
imaging module.
Indirect illumination:
• For observation in regredient light (indirect illumination), the centring screw (18)
is loosened in order to move the slit image out of the centre of the visual field.
Tightening the screw centres the slit image again.
Slit tilting:
• The latch (36) can be used to tilt the illumination in 5° steps. This creates an
angled light beam during horizontal slit orientation. Tilting the slit enables reflex-
free examination with contact lenses (fundus and gonioscopy) and magnifying
lenses.