6.4 Setting the filters & diaphragms
a. Open
b. Grey filter (10%)
c. Red removal filter
d. Reserve opening for filter of choice ø 15 mm (0 / –0.2), thickness 2.5 mm
e. Fixation star (predominantly used for fixation examination of cross-eyed
children with amblyopia)
f. Apertures of ø 8, 5, 3, 2, 1 and 0.2 mm
g. Display of slit length adjustment from 1 to 14 mm
h. Blue filter
6.5 Fixation star
• Turning the diaphragm disc as far as it will go to the left switches
on the fixation star and the symbol "S" appears in the window. In
some examinations of the fundus, this star is projected onto the
ocular fundus and is also visible to the patient, who is asked to
focus on the centre hole of the star. This allows the examiner to
see where the patient's vision is most focused.
• A typical use of the fixation star is during laser treatments close to
the macula. Similarly, it is also possible to identify
microstrabismus with the projection of the fixation star. The
fixation star is usually used with an upstream red removal filter.
6.6 Microscope and eyepiece
44. Frontal objective f 103 mm
45. Rotary knob displaying the set magnification
46. 3-level magnification changer (Galilei
system)
47. Quick-release fastener for knurled ring
48. Binocular tube f 130 mm with convergent
insight, pupil distance can be set from 52 −
78 mm
49. Eyepiece 12.5× / field of vision ø 16 mm
50. Index (white point)
51. Knurled ring with dioptre scale for setting
the refraction of the examiner (± 6 D)
52. Sliding occluder (for people who wear
glasses)
44 45 46
47 48 49 50
52
51
7 Decommissioning
The LED illumination can be switched off with the illumination controls. The power
supply remains switched on during this process and the switch lights up green. To
switch the system off completely, the rocker switch must be set to the 0 = 'OFF'
position. This creates a two-pole isolation from the mains.
NOTE!
Disconnect the power supply from the mains if you do not intend to use
it for an extended period of time.