F.  Information and manufacturer's declaration  

 

concerning electromagnetic compatibility  

 (EMC)

F.1 General

The BM 900 slit lamp system fulfills the requirements on electromagnetic compati

-

bility according to EN 60601-1-2:2007 (IEC 3. Edition) + EN 60601-1-2:2015 (IEC 4. 

Edition). The instrument is built so that the generation and emission of electromag-

netic interference is limited to the extent that other devices are not disturbed in their 

use in accordance with the regulations and so that the instrument itself is suitably 

immune to electromagnetic interference.

WARNING!

• 

Electrical medical devices and systems are subject to special EMC 

measures and must be installed in accordance with the EMC instruc-

tions contained in this accompanying document.

The operation of other lines or equipment than those listed may lead 

to higher emissions or may reduce the device's resistance to interfe-

rence.

Third-party devices may only be connected in compliance with the 

EN 60601-1 standard.

 

 

F.2  Emitted interference (standard table 1)

The information is based on the requirements of EN 60601-1-2:2007 (IEC 3rd edition) and EN 60601-1-2:2015 (IEC 4

th

 edition).

Guidance and manufacturer's declaration – electromagnetic emissions
This product is intended for use in the electromagnetic environment specified below. The customer or the user of this product should assure that it is used in such an envi-

ronment

RF emissions CISPR 11

Group 1

This product uses RF energy only for its internal function. Therefore, its RF emissions are very low and 

are not likely to cause any interference in nearby electronic equipment.

RF emissions CISPR 11

Class B

This product is suitable for use in all establishments, including domestic establishments and those 

directly connected to the public low-voltage power supply network that supplies buildings used for 

domestic purposes.

Emission of harmonics according to 

EN 61000-3-2

Class A

Voltage fluctuations / flicker emissions 

according to EN 61000-3-3

Fulfilled

Emission test

Compliance

Electromagnetic environment - guidance

01_IFU_BM900-7200492-04150_eng.indd   12

21.02.2017   09:55:00

Содержание BM 900

Страница 1: ...00 7200492 04150 15 Edition 2017 01 DEUTSCH ENGLISH FRAN AIS ITALIANO ESPA OL NEDERLANDS PORTUGU S SVENSKA DOC no 1500 1500 1400209 04000 INSTRUCTIONS FOR USE Slit lamp BM 900 15 Edition 2017 01 01_IF...

Страница 2: ...of diseases or trauma which affect the structural properties of the eye Contraindication There is no absolute contraindication for tests with this device Appropriate profes sional judgement and cautio...

Страница 3: ...Preparing the patient 7 5 3 Operating the instrument 7 5 4 Setting the filters diaphragms 8 5 5 Fixation star 9 5 6 Microscope and eyepiece 9 6 Decommissioning 9 7 Technical data 9 7 1 Slit illuminat...

Страница 4: ...npack the appliance check whether the packaging shows traces of incorrect handling or damage If this is the case notify the transport company that has delivered the goods to you Unpack the equipment t...

Страница 5: ...nued use of a product damaged by incorrect handling may lead to personal injury In such a case the manufacturer will not accept any liability Haag Streit does not grant any warranties either expressed...

Страница 6: ...Breath shield option 14 Protective cover 15 Illumination unit microscope an gle scale 16 Illumination arm locking screw 17 Microscope arm locking screw 18 Slit width setting screw 19 Slit length diap...

Страница 7: ...s and press the rocker switch The green light on the rocker switch illuminates when the device is switched on Turn the rotating knob on the illumination control to a position between 1 and 10 5 Operat...

Страница 8: ...d 10 a short mirror 12 is used the illumination turned 90 using the locking screw 20 and tilted in 5 steps us ing the latch 27 and the illumination and microscope turned to the central po sition 0 The...

Страница 9: ...Dioptre scale for 16x eyepieces for setting the refraction of the examin er 8 D 46 Dioptre scale for 10x eyepieces for setting the refraction of the examin er 8 D 47 Index white point 48 10x or 25x e...

Страница 10: ...from the to the side The structure of the test rod must be in sharp focus in all magnifications Close slit edges to approx 0 5 mm The borders must be in sharp focus here Completely open slit edges and...

Страница 11: ...15004 2 standards into account The EN 60601 1 standard must be observed when using different medical and or non medical electrical devices in combination Compliance of the BM 900 slit lamp with the Di...

Страница 12: ...sistance to interfe rence Third party devices may only be connected in compliance with the EN 60601 1 standard F 2 Emitted interference standard table 1 The information is based on the requirements of...

Страница 13: ...or hospital environment Surge EN 61000 4 5 1 kV for symmetrical voltages 2 kV for asymmetrical voltages 1 kV for symmetrical voltages 2 kV for asymmetrical voltages Mains power quality should be that...

Страница 14: ...ded separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interf...

Страница 15: ...minal output of the transmitter W Safe distance according to transmission frequency m 150 kHz 80 MHz D 0 35 80 MHz 800 MHz D 0 7 800 MHz 2 7 GHz D 1 4 0 01 0 1 1 10 100 For transmitters with a nominal...

Страница 16: ...Should you have any further questions please contact your HAAG STREIT dealer at http www haag streit com contact contact your distributor html HAAG STREIT AG Gartenstadtstrasse 10 3098 Koeniz Switzerl...

Отзывы: