Total magnification

10x / 16x / 25x / 40x

Object field diameter

18 mm / 11.3 mm / 8 mm / 5 mm

7.3  Instrument base

Operation

Single-handed operation of control lever in three 

dimensions 

Adjustment of instrument base 100 mm (length), 100 mm (side) , 30 mm (height)

7.4 Dimensions

Weight:

11.4 kg (without power supply, head rest and options)

Dimensions L x W x H: 332 x 290 x 700 mm
Packaging L x W x H:  420 x 510 x 780 mm

 

8. Maintenance

WARNING!

 

Do not modify this equipment without authorization of the manufacturer.

Installation and repairs may only be performed by trained specialists. 

Contact your HAAG-STREIT representative for installation, repairs and 

modification work on the system. The contact details are available at 

www.haag-streit.com.

Only original HS replacement parts may be used.

 

 

The LED illumination can be operated maintenance-free for its entire service life.

8.1  Device inspection 

In order to correctly check the slit lamp, proceed as follows:

Insert test rod into the radial movement bearing, whilst at the same time aligning

   the surface to the microscope at a right angle

Set slit length to 8 mm 

Set llumination strength to 50%

Set magnification to max. in the microscope.

Set the eyepieces in such a way that the test rod is in sharp focus. Turn the 

  eyepiece from the (+) to the (-) side.

The structure of the test rod must be in sharp focus in all magnifications.

Close slit edges to approx. 0.5 mm. The borders must be in sharp focus here.

Completely open slit edges and turn the test rod by 45°, the sharp area must 

  be in the centre of the test rod.

8.2 Repair

To guarantee a long service life, the device must be cleaned weekly as described 

and protected with the dust cover when not in use. We recommend having the de-

vice inspected once a year by an authorized service technician.

8.3 Cleaning

• 

Only clean the housing with a dry cloth.

• 

Do not use any liquids, alcohol or corrosive substances.

• 

The exposed glass surfaces can be dusted with a dusting brush. 

NOTE!

IP code: IPX0 (device is not protected against liquids)

8.4  Replacing the illumination mirror

The mirror can be most easily accessed if the microscope is turned away from the

illumination and the illumination inclined two points.

WARNING!

Only use mirrors with a LOT number.

8.5  Dust cover

We recommend protecting the slit lamp with a dust cover 

(53) when not in use. 

A. Appendix

A.1  Accessories / spare parts

53

WARNING!

 

Do not modify this equipment without authorization of the manufacturer.

Installation and repairs may only be performed by trained specialists. 

Contact your HAAG-STREIT representative for installation, repairs and 

modification work on the system. The contact details are available at 

www.haag-streit.com.

Only original HS replacement parts may be used.

 

 

NOTE!

An asterisk (*) shows that you should contact your HAAG-STREIT rep-

resentative for further information.

01_IFU_BM900-7200492-04150_eng.indd   10

21.02.2017   09:54:59

Содержание BM 900

Страница 1: ...00 7200492 04150 15 Edition 2017 01 DEUTSCH ENGLISH FRAN AIS ITALIANO ESPA OL NEDERLANDS PORTUGU S SVENSKA DOC no 1500 1500 1400209 04000 INSTRUCTIONS FOR USE Slit lamp BM 900 15 Edition 2017 01 01_IF...

Страница 2: ...of diseases or trauma which affect the structural properties of the eye Contraindication There is no absolute contraindication for tests with this device Appropriate profes sional judgement and cautio...

Страница 3: ...Preparing the patient 7 5 3 Operating the instrument 7 5 4 Setting the filters diaphragms 8 5 5 Fixation star 9 5 6 Microscope and eyepiece 9 6 Decommissioning 9 7 Technical data 9 7 1 Slit illuminat...

Страница 4: ...npack the appliance check whether the packaging shows traces of incorrect handling or damage If this is the case notify the transport company that has delivered the goods to you Unpack the equipment t...

Страница 5: ...nued use of a product damaged by incorrect handling may lead to personal injury In such a case the manufacturer will not accept any liability Haag Streit does not grant any warranties either expressed...

Страница 6: ...Breath shield option 14 Protective cover 15 Illumination unit microscope an gle scale 16 Illumination arm locking screw 17 Microscope arm locking screw 18 Slit width setting screw 19 Slit length diap...

Страница 7: ...s and press the rocker switch The green light on the rocker switch illuminates when the device is switched on Turn the rotating knob on the illumination control to a position between 1 and 10 5 Operat...

Страница 8: ...d 10 a short mirror 12 is used the illumination turned 90 using the locking screw 20 and tilted in 5 steps us ing the latch 27 and the illumination and microscope turned to the central po sition 0 The...

Страница 9: ...Dioptre scale for 16x eyepieces for setting the refraction of the examin er 8 D 46 Dioptre scale for 10x eyepieces for setting the refraction of the examin er 8 D 47 Index white point 48 10x or 25x e...

Страница 10: ...from the to the side The structure of the test rod must be in sharp focus in all magnifications Close slit edges to approx 0 5 mm The borders must be in sharp focus here Completely open slit edges and...

Страница 11: ...15004 2 standards into account The EN 60601 1 standard must be observed when using different medical and or non medical electrical devices in combination Compliance of the BM 900 slit lamp with the Di...

Страница 12: ...sistance to interfe rence Third party devices may only be connected in compliance with the EN 60601 1 standard F 2 Emitted interference standard table 1 The information is based on the requirements of...

Страница 13: ...or hospital environment Surge EN 61000 4 5 1 kV for symmetrical voltages 2 kV for asymmetrical voltages 1 kV for symmetrical voltages 2 kV for asymmetrical voltages Mains power quality should be that...

Страница 14: ...ded separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interf...

Страница 15: ...minal output of the transmitter W Safe distance according to transmission frequency m 150 kHz 80 MHz D 0 35 80 MHz 800 MHz D 0 7 800 MHz 2 7 GHz D 1 4 0 01 0 1 1 10 100 For transmitters with a nominal...

Страница 16: ...Should you have any further questions please contact your HAAG STREIT dealer at http www haag streit com contact contact your distributor html HAAG STREIT AG Gartenstadtstrasse 10 3098 Koeniz Switzerl...

Отзывы: