13.11 Interfaces (RS232, D-SUB, 9-pole)
Description
Data
COM 1 (standard)
Baud rate 9600 bps / 115200 bps
COM 2 (standard)
Baud rate 9600 bps
13.12 Voltage, Current, Performance
Description
Data
Input voltage *
400 V AC 3 ~ / N / PB (50/60 Hz)
Current input (long time) *
12.1 A
Current input (momentary) *
22.5 A
Energy consumption (long time)
≤ 4844 VA
Energy consumption (momentary)
≤ 8996 VA
Energy efficiency
N/A
Device protection switch (circuit breaker)
32 A
Drive motor capacity
11000 W
Elevation motor capacity
370 W
Device leakage current
≤ 0.2 mA
Isolation transformer
9000 VA
Power supply cord
detachable, 3 m
* Overload or weak power supply may lead to reduced speed accuracy or tripping the fuse.
Subject to technical alterations without prior notice. E & OE (errors and omissions excepted).
Note: In case of special voltage versions, the data on the name plate apply.
Connect mains only if data on nameplate and local power supply are identical.
13.13 Software, Programs
Description
Data
h/p/cosmos para control
®
included
h/p/cosmos para graphics
®
(non medical)
optional
h/p/cosmos para analysis
®
(non medical)
optional
h/p/cosmos para motion
®
(non medical)
optional
gaitway (non medical)
optional
Noraxon (non medical)
optional
Microgate Optogait
optional
zebris Rehawalk
optional
Number of stored programmes
6 training profiles,
10 test profiles,
8 user definable profiles
For other coscom compatible software see website
www.coscom.org
13.14 All-pole disconnection
The following options are available for all-pole disconnection:
❚
Unplug device from power socket.
❚
Unplug cable from device (if possible).
❚
Switch off device protection switch
Maintain enough free space to ensure access to cables and the circuit breaker (see “position of subject and operator”).
med