page 17 of 90
cos105000_200-300_75-125, rev. 1.03
© 1988 - 2019 h/p/cosmos sports & medical gmbh
❚
Do not turn around, walk sideways or backwards; do not jump on or off the running belt while it is in motion.
❚
Do not touch the running belt while it is in motion (besides contact with feet).
❚
Do not lean on the UserTerminal - do not apply pressure to the displays - press keys softly.
❚
Ensure assist mean, accessories, cables etc. do not extend into the running area.
❚
Do not insert any object (especially no metal objects such as a pin or a wire) into any gap or any outlet on the device.
❚
Do not touch the subject and external electrical devices at the same time.
❚
the four modes. Only stop command has higher priority and cannot be overwritten.
❚
Be aware that electromagnetic interferences may cause a fail-safe mode, the running belt will stop with a pre-defined deceleration
ramp.
❚
WARNING: To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective earth.
❚
WARNING: Do not use portable high frequency communication devices in the subject environment (see “position of subject
and user”). Disregard can cause loss of performance.
❚
Free standing equipment has to be installed on a stable and levelled base.
❚
Choose proper floor, clothing and humidity, in order to prevent electrostatic discharge (also see technical data).
❚
Do not use the device without instruction by authorized personnel acc. to the instruction protocol.
❚
Regard safety area behind device of 2.0 m x width of treadmill.
❚
❚
maintenance and training in order to prevent being captured in trapping zones.
❚
Perform a daily visual inspection (see chapter “maintenance”).
❚
Obey the maintenance intervals claimed in chapter “maintenance”.
❚
Obey the competences claimed in chapter “maintenance”.
❚
A second person has to be present during maintenance.
❚
reconnection, mark clearly and inform h/p/cosmos service personnel via telephone and writing.
❚
In case of any visible or assumed wear and tear (of the device, accessories, labels, etc.), unplug device, exclude reconnection,
mark clearly and inform h/p/cosmos service personnel via telephone and writing. Do not change or remove any labels!
❚
personnel via telephone and writing.
❚
Do not modify the device, configurations, accessories or software in any way.
❚
❚
Disinfect the device before and after every treatment.
❚
Disconnect the device and all accessories from mains power supply before cleaning or disinfection.
med