page 54 of 90
cos105000_200-300_75-125, rev. 1.03
© 1988 - 2019 h/p/cosmos sports & medical gmbh
12.3 Error messages
Following error messages may be displayed on the control panel.
Error code
Acoustic code
(x = short beep,
o = long beep)
Error message
Action
E01
ooooo xoooo
Oil Help
See “lubrication”
E02
ooooo xxooo
Service Help
See “safety inspection”
E20
xxooo ooooo
Elev Help
Unplug device,
exclude reconnection and
contact
E21
xxooo xoooo
Incr Help
E30
xxxoo ooooo
Setup Help
E31
xxxoo xoooo
Setup Help
E32
xxxoo xxooo
Setup Help
E41
xxxxo xoooo
Setup Help
E50
xxxxx ooooo
FU Help
E51
xxxxx xoooo
FU Help
E52
xxxxx xxooo
FU Help
Description
Action
Error message
elevation speed
Switch for slow or fast change of
elevation
(fast: switch is blue lighted)
belt direction
Switch for forward / backward
running
(backward: switch is blue lighted)
Belt alignment – adjust it! Running belt is moving outside
tolerance range.
lighted: belt out of tolerance
blinking: belt severe out of tolerance
check power supply!
Contact
[email protected].
oil tank empty
See 11.3 Lubrication
Emergency off /
emergency stop activated
Check Emergency off / emergency
stop (see “safety”).
In case there is no improvement,
contact [email protected].
Switch on delay
For protection of electric components
the green indicator is blinking for
1min. until device can be started
(indicator lighted green)
On – Button
Switches the device on (only
possible after switch on delay)
Off – Button
Switches the device off (green
indicator is blinking for 1min)
Following error messages may be displayed on the UserTerminal of the control panel.